Cook Inc.: VentriClear II Ventricular Drainage Catheter Set. Allows external access and drainage of cerebrospinal fluid (CSF) from the ventricles of the brain.
- Recall number
- Z-0416-2016
- Recalling firm
- Cook Inc.
- Product
- VentriClear II Ventricular Drainage Catheter Set. Allows external access and drainage of cerebrospinal fluid (CSF) from the ventricles of the brain.
- Status
- Terminated
- Initiated
- 2015-11-19
- Posted
- 2015-12-14
- Terminated
- 2017-02-22
- Reason
- Cook Medical initiated a voluntary recall of specific lot numbers of the VentriClear II Ventricular Drainage Catheter Set due to the device being sterilized twice, which may result in the minocycline and rifampin catheter coating to be less than what is indicated on the labeling.
- Root cause
- Process control
- Action
- Cook Medical issued letter via certified mail on 11/19/2015, for their voluntary recall of the VentriClear II Ventricular Drainage Catheter Set due to being sterilized twice. The Distributor was instructed to forward the recall letter on to its customers, and quarantine any unused recalled devices, and return them to Cook Medical. Customers with questions may contact: Cook Medical Customer Relations 1-800-457-4500 or 1-812-339-2235. Monday through Friday between 7:30a.m. and 5:00p.m. Eastern Daylight Time or email at CustomerRelationsNA@cookmedical.com.
- Quantity
- 35
- Distribution
- Distributed in Tennessee, Florida, Michigan, and Pennsylvania and in Saudi Arabia.
- Lot, serial or model codes
- Catalog Number: 50318Lot Numbers: F4365351X, F4692818X, NS5274254X, NS5667436X, NS5727900X, NS6079165X, NS6079166X
- 510(k) numbers
- ["K071640"]
- PMA numbers
- not on file
- Product code
- NHC
- Firm FEI
- 1820334
- Firm city
- Bloomington
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28