Skip to the content

FDA Medical Device Recalls (openFDA)

Cook Inc.: VentriClear II Ventricular Drainage Catheter Set. Allows external access and drainage of cerebrospinal fluid (CSF) from the ventricles of the brain.

Recall number
Z-0416-2016
Recalling firm
Cook Inc.
Product
VentriClear II Ventricular Drainage Catheter Set. Allows external access and drainage of cerebrospinal fluid (CSF) from the ventricles of the brain.
Status
Terminated
Initiated
2015-11-19
Posted
2015-12-14
Terminated
2017-02-22
Reason
Cook Medical initiated a voluntary recall of specific lot numbers of the VentriClear II Ventricular Drainage Catheter Set due to the device being sterilized twice, which may result in the minocycline and rifampin catheter coating to be less than what is indicated on the labeling.
Root cause
Process control
Action
Cook Medical issued letter via certified mail on 11/19/2015, for their voluntary recall of the VentriClear II Ventricular Drainage Catheter Set due to being sterilized twice. The Distributor was instructed to forward the recall letter on to its customers, and quarantine any unused recalled devices, and return them to Cook Medical. Customers with questions may contact: Cook Medical Customer Relations 1-800-457-4500 or 1-812-339-2235. Monday through Friday between 7:30a.m. and 5:00p.m. Eastern Daylight Time or email at CustomerRelationsNA@cookmedical.com.
Quantity
35
Distribution
Distributed in Tennessee, Florida, Michigan, and Pennsylvania and in Saudi Arabia.
Lot, serial or model codes
Catalog Number: 50318Lot Numbers: F4365351X, F4692818X, NS5274254X, NS5667436X, NS5727900X, NS6079165X, NS6079166X
510(k) numbers
["K071640"]
PMA numbers
not on file
Product code
NHC
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register