Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: The Trilogy Acetabular System Shell with Cluster Holes Porous 56 mm O.D. Sterile The Trilogy Acetabular System is indicated for either cemented or uncemented use In skeletally mature Individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused Hip, fracture of the pelvis, and diastrophic variant. The shells with screw holes permit the use of titanium alloy screws to provide additional fixation and security, particularly in those cases where acetabular bone stock is deficient.

Recall number
Z-0416-2013
Recalling firm
Zimmer, Inc.
Product
The Trilogy Acetabular System Shell with Cluster Holes Porous 56 mm O.D. Sterile The Trilogy Acetabular System is indicated for either cemented or uncemented use In skeletally mature Individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused Hip, fracture of the pelvis, and diastrophic variant. The shells with screw holes permit the use of titanium alloy screws to provide additional fixation and security, particularly in those cases where acetabular bone stock is deficient.
Status
Terminated
Initiated
2012-09-20
Posted
2012-11-21
Terminated
2014-04-18
Reason
Zimmer Trilogy Acetabular System Shell with Cluster Holes, Porous,56 MM O.D., sterile; is missing polar hole threads. Zimmer received one complaint in which no injury was reported.
Root cause
Process control
Action
Zimmer sent a Urgent Medical Device Recall Notice dated September 20, 2012, to all affected customers. The letter identified the products the problem and the action needed to be taken by the customer. Customers were instructed to locate quarantine and return affected product to Zimmer Product Service Department, 1777 West center St. Warsaw, IN 46580 for product in the USA. Customers were asked to complete the Inventory Return Certification Form and follow the instructions where to return this form. For further questions please call (574) 372-4807.
Quantity
24 USA; 71 worldwide
Distribution
Worldwide Distribution--USA (nationwide) including the states of AL, CA, CT, KS, MA, MD, MN, MO, NY, SD, TX and WI and the countries of Bulgaria,Canada, Germany, Italy, Netherlands, Russian Fed, Sweden and UK
Lot, serial or model codes
REF 6200-056-22, lot 62010615
510(k) numbers
not on file
PMA numbers
not on file
Product code
LPH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register