Skip to the content

FDA Medical Device Recalls (openFDA)

Acclarent, Inc.: TruDi NAV Suction Instruments: 0-Degree, REF: TDNS000Z; 70-Degree, REF: TDNS070Z; and 90-Degree, REF: TDNS090Z

Recall number
Z-0415-2022
Recalling firm
Acclarent, Inc.
Product
TruDi NAV Suction Instruments: 0-Degree, REF: TDNS000Z; 70-Degree, REF: TDNS070Z; and 90-Degree, REF: TDNS090Z
Status
Open, Classified
Initiated
2021-11-12
Posted
not on file
Terminated
not on file
Reason
Suction Instruments were incorrectly calibrated so they may not meet the system accuracy specification for visual verification of device location in the patient anatomy. If the surgeon does not detect the accuracy error, this could lead to an intraprocedural delay, a potential for reversible tissue damage or damage to the carotid artery, CSF leak, or damage to optic nerve.
Root cause
Process control
Action
On 11/12/21, recall notices were sent to OR Directors, Risk Managers, or Material Managers. Customers were asked to do the following: 1) Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2) Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3) If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4) Complete the Business Reply Form and return to acclarent8930@sedgwick.com 5) After the Business reply Form is received a return kit will be sent. 6) Keep this notice visibly posted for awareness until all product subject to this recall has been returned. Customers with additional questions are encouraged to call the recalling firm at 1-877-775-2789, 4:00am-5:00pm PST.
Quantity
1543
Distribution
US: AZ, MI, IA, LA, IL, VA, SC, NY, IN, OR, SD, TX, PA, NE, FL, CT, MN, NV, CA, NJ, UT, AL, MA, WI, TN, WV, OH, NC, KY, HI¿, OK, GA, CO, MO, WY, MS, NH, HI, MD, ID
Lot, serial or model codes
0-Degree Lots: 1903120, 1903250, 1904060, 1904180, 1904240, 1905030, 1905070, 1906260, 1907120, 1907170, 1907220, 1907240, 1907260, 2003040, 2009110, 2011240, 2102080, 2103090, 2104260, 2105040, 2105260, 2106100, 2107070. 70-Degree Lots: 1903137, 1903257, 1904067, 1905017, 1905037, 1905077, 1906067, 1906267, 1907127, 1908137, 1908157, 2003127. 90-Degree: 1903139, 1903259, 1904069, 1905019, 1905139, 1906079, 1908269, 1909279, 1909309, 1910299, 2003029, 2004029
510(k) numbers
["K180948"]
PMA numbers
not on file
Product code
PGW
Firm FEI
3005172759
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register