Abbott Laboratories, Inc: ARCHITECT i2000 Processing Module, for in vitro diagnostics, List Number 08C89-01; Temperature Controller Board Part #78560-109; distributed by Abbott Laboratories, Irving, TX.

Recall number
Z-0415-2007
Recalling firm
Abbott Laboratories, Inc
Product
ARCHITECT i2000 Processing Module, for in vitro diagnostics, List Number 08C89-01; Temperature Controller Board Part #78560-109; distributed by Abbott Laboratories, Irving, TX.
Status
Terminated
Initiated
2006-12-12
Posted
2007-02-07
Terminated
2007-09-24
Reason
The re-designed ARCHITECT Temperature Controller Board has been identified as being susceptible to electromagnetic interference (EMI) in the laboratory. This can cause a board reset condition, stopping temperature control function of the board. Situation does not stop analyzer operation and does not generate an error condition alerting operator when condition occurs.
Root cause
Other
Action
Firm contacted consignees by telephone and fax with instructions to verify proper functioning of the Temperature Controller Board starting December 13, 2006.
Quantity
57 units.
Distribution
Worldwide, including USA, Argentina, Australia, Brazil, Canada, Chile, Columbia, Dominican Republic, Germany, Hong Kong, Jamaica, Japan, Korea, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Taiwan, Thailand, Trinidad, Tobago, Uruguay, and Venezuela.
Lot, serial or model codes
Serial numbers: 1202790, 1202804, 1202805, 1202806, 1202807, 1202808, 1202809, 1202810, 1202811, 1202812, 1202813, 1202814, 1202815, 1202816, 1202817, 1202818, 1202819, 1202820, 1202821, 1202822, 1202823, 1202824, 1202825, 1202826, 1202827, 1202828, 1202829, 1202830, 1202831, 1202832, 1202833, 1202834, 1202835, 1202836, 1202837, 1202838, 1202839, 1202840, 1202841, 1202842, 1202843, 1202844, 1202845, 1202846, 1202847, 1202848, 1202849, 1202850, 1202851, 1202852, 1202853, 1202854, 1202855, 1202856, 1202857, 1202858, 1202859.
510(k) numbers
["K983212"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1628664
Firm city
Irving
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register