Skip to the content

FDA Medical Device Recalls (openFDA)

Cook Inc.: Skinny Needle with Chiba Tip. RPN 090010, 090010-ET; GPN G14254, G15290. Rigid needle available with EchoTip to provide enhanced visualization of needle tip when used with ultrasonic imaging equipment.

Recall number
Z-0414-2022
Recalling firm
Cook Inc.
Product
Skinny Needle with Chiba Tip. RPN 090010, 090010-ET; GPN G14254, G15290. Rigid needle available with EchoTip to provide enhanced visualization of needle tip when used with ultrasonic imaging equipment.
Status
Open, Classified
Initiated
2021-10-13
Posted
not on file
Terminated
not on file
Reason
Specific lots may contain black debris on the exterior of the inner stylet and/or on the interior of the outer cannula. Local inflammatory reactions may occur if an affected product is used.
Root cause
Process change control
Action
On October 13, 2021, the firm distributed Urgent Medical Device Recall letters to customers. Customers were informed that the needles may contain black debris on the exterior of the inner stylet and/or the interior of the outer cannula. Potential adverse events that may occur if an affected product is used include local/self-limited inflammatory reactions. To date, Cook has received no complaints related to this issue. However, please be advised that the presence of black debris may go undetected by the user. Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of these products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Action being Taken by Cook Medical Cook Medical is removing potentially impacted devices from the market. Additionally, an investigation has been opened to determine the appropriate corrective action(s) to prevent reoccurrence of a similar issue. If you have any questions or concerns, please contact Cook Medi
Quantity
112 US, 240 OUS
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CA, IN, LA, MS, MT, NC, NY, OR, PA, WA and the countries of France, Italy, and Mexico.
Lot, serial or model codes
RPN 090010/GPN G14254: LOT NUMBER UDI 13977925 (01)00827002142541(17)240524(10)13977925 14030949 (01)00827002142541(17)240617(10)14030949 14054162 (01)00827002142541(17)240628(10)14054162 14116047 (01)00827002142541(17)240728(10)14116047 RPN 090010-ET/GPN G15290: LOT NUMBER UDI 14038615 (01)00827002152901(17)240618(10)14038615
510(k) numbers
not on file
PMA numbers
not on file
Product code
LJE
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register