Steris Corp: STERIS Harmony LL 500/700 Surgical Light, Vertical Spring-Loaded Arm. Intended for local illumination of the operating and examination area of the patient in treatment rooms.

Recall number
Z-0414-2010
Recalling firm
Steris Corp
Product
STERIS Harmony LL 500/700 Surgical Light, Vertical Spring-Loaded Arm. Intended for local illumination of the operating and examination area of the patient in treatment rooms.
Status
Terminated
Initiated
2009-09-11
Posted
2009-11-23
Terminated
2012-03-01
Reason
Drger Medical AG & Co., KGaA informed STERIS Corporation that they received (4) reported international cases where a welded seam at the joint of the spring-loaded arm of the Drger SOLA 500/700 surgical light broke as a result of a crack that had slowly developed over time. The STERIS Harmony LL 500/700 surgical light is a Private Label version of the Drger SOLA 500/700 surgical light.
Root cause
Device Design
Action
The firm sent Urgent Product Correction letters, dated Sept. 11, 2009, to their customers. The letter identified the problem and stated that Drager had supplied STERIS the replacement spring arms and that a STERIS Service Technician would perform the replacement. The letter also stated that the correction would be performed at no charge to the customers. According to the letter, customers would be contacted by their STERIS Service Representative to arrange a convenient time for replacement of potentially affected spring arms. In the meantime, the letter states that customers should use their surgical lights in accordance with the guidelines presented in the Operator's Manual. Questions or concerns regarding the letter should be directed to Barbara Rich at 440-392-7318 or your local STERIS account manager.
Quantity
388 units
Distribution
Worldwide Distribution -- USA, including the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MA, MN, MO, MS, NC, NH, NJ, NV, NY, OH, PA, RI, SC, TN, TX, UT, VA, WA, WI & WY and the country of Canada.
Lot, serial or model codes
Part Number: YG18568; Serial Numbers: Year 2002 - 0111 072 XXXXX to 0111 122 XXXXX; Year 2003-0111 013 XXXXX to 0111 123 XXXXX; Year 2004-0111 014 XXXXX to 0111 124 XXXXX; Year 2005-0111 015 XXXXX to 0111 125 XXXXX; and Year 2006-0111 016 XXXXX to 0111 066 XXXXX - (XXXXX means every 5-digit serial number).
510(k) numbers
["K010724"]
PMA numbers
not on file
Product code
FTD
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register