Steris Corp: STERIS Harmony LL 500/700 Surgical Light, Vertical Spring-Loaded Arm. Intended for local illumination of the operating and examination area of the patient in treatment rooms.
- Recall number
- Z-0414-2010
- Recalling firm
- Steris Corp
- Product
- STERIS Harmony LL 500/700 Surgical Light, Vertical Spring-Loaded Arm. Intended for local illumination of the operating and examination area of the patient in treatment rooms.
- Status
- Terminated
- Initiated
- 2009-09-11
- Posted
- 2009-11-23
- Terminated
- 2012-03-01
- Reason
- Drger Medical AG & Co., KGaA informed STERIS Corporation that they received (4) reported international cases where a welded seam at the joint of the spring-loaded arm of the Drger SOLA 500/700 surgical light broke as a result of a crack that had slowly developed over time. The STERIS Harmony LL 500/700 surgical light is a Private Label version of the Drger SOLA 500/700 surgical light.
- Root cause
- Device Design
- Action
- The firm sent Urgent Product Correction letters, dated Sept. 11, 2009, to their customers. The letter identified the problem and stated that Drager had supplied STERIS the replacement spring arms and that a STERIS Service Technician would perform the replacement. The letter also stated that the correction would be performed at no charge to the customers. According to the letter, customers would be contacted by their STERIS Service Representative to arrange a convenient time for replacement of potentially affected spring arms. In the meantime, the letter states that customers should use their surgical lights in accordance with the guidelines presented in the Operator's Manual. Questions or concerns regarding the letter should be directed to Barbara Rich at 440-392-7318 or your local STERIS account manager.
- Quantity
- 388 units
- Distribution
- Worldwide Distribution -- USA, including the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MA, MN, MO, MS, NC, NH, NJ, NV, NY, OH, PA, RI, SC, TN, TX, UT, VA, WA, WI & WY and the country of Canada.
- Lot, serial or model codes
- Part Number: YG18568; Serial Numbers: Year 2002 - 0111 072 XXXXX to 0111 122 XXXXX; Year 2003-0111 013 XXXXX to 0111 123 XXXXX; Year 2004-0111 014 XXXXX to 0111 124 XXXXX; Year 2005-0111 015 XXXXX to 0111 125 XXXXX; and Year 2006-0111 016 XXXXX to 0111 066 XXXXX - (XXXXX means every 5-digit serial number).
- 510(k) numbers
- ["K010724"]
- PMA numbers
- not on file
- Product code
- FTD
- Firm FEI
- 1527821
- Firm city
- Mentor
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28