FDA Medical Device Recalls (openFDA)
Randox Laboratories, Limited: Randox Liquid Cardiac Controls Catalogue Number CQ5052.
- Recall number
- Z-0413-2020
- Recalling firm
- Randox Laboratories, Limited
- Product
- Randox Liquid Cardiac Controls Catalogue Number CQ5052.
- Status
- Terminated
- Initiated
- 2019-10-04
- Posted
- not on file
- Terminated
- 2020-07-06
- Reason
- Removal of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5052 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.
- Root cause
- Under Investigation by firm
- Action
- On October 1, 2019 the firm distributed Urgent Medical Device Recall letters by e-mail instructing them to: - Inspect your stock and quarantine affected stock. - Replace the value sheet in the kit with the revised value sheet provided. - Randox is not recommending a review of previous results as changes in quality control recovery would be reviewed at the time of occurrence. - Discuss the contents of this notice with your Medical Director. - Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization.
- Quantity
- 7,505
- Distribution
- US Distribution: WV, CA, DE, NJ, NY, GA, IN, PR OUS Distribution: CA, PL, CH, JP, NL, IT, CN, GB, IE, RU, MD, BY, SA, VN, AE, BH, DK, CO, ES, FR, TW, PH, TH, OM, HU, IN, KR, ZA, AU, CZ, PE, CL, IR, NZ, BE, DE, BA, SV, SK, HK, PT, PK, BR, TR, UA, SE, QA, LT, PE,
- Lot, serial or model codes
- Catalogue Number: CQ5052 GTIN: 05055273207453 Batch/lot: 4244CK (Exp. 28Nov2019), 4247CK (Exp. 28Nov2019), 4261CK (Exp. 28Nov2019), 4312CK Exp. 28Jun2020) and 4315CK (Exp. 28Jun2020)
- 510(k) numbers
- ["K110534"]
- PMA numbers
- not on file
- Product code
- JJY
- Firm FEI
- 1000361607
- Firm city
- Crumlin
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28