Philips Medical Systems North America Co. Phillips: Philips EasyVision MM workstation with software R10.2 and R11.1 The EasyVision MM Workstation is intended for the manipulation and display of medical images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. Device options enable mammography reading, teleradiology, fast demonstration, 3D, etc. Device accessories enable orthopedic pre-operative planning; angiography/cardiology imaging capabilities; and bone densitometry capabilities. Typical users are trained professionals such as physicians, radiologists, nurses, medical technicians and assistants.

Recall number
Z-0412-2011
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Philips EasyVision MM workstation with software R10.2 and R11.1 The EasyVision MM Workstation is intended for the manipulation and display of medical images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. Device options enable mammography reading, teleradiology, fast demonstration, 3D, etc. Device accessories enable orthopedic pre-operative planning; angiography/cardiology imaging capabilities; and bone densitometry capabilities. Typical users are trained professionals such as physicians, radiologists, nurses, medical technicians and assistants.
Status
Terminated
Initiated
2009-01-30
Posted
2010-11-18
Terminated
2010-11-19
Reason
Review by firm of Philips EasyVision MM workstations found that to assure compliance with 21 CFR 900.12(c)(5) corrected software was needed so that all sites are in compliance.
Root cause
Other
Action
Philips Medical Systems North America Co. sent an URGENT-Field Safety Notice dated January 30, 2009, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were advised to ascertain the quality of a hard or soft copy explicity before further distribution to make sure that no part of the tissue is covered by image identification information. Philips will issue the Field Change Order (FCO) 83000129 and release a level on EasyVision MM R10.2 L6 and EasyVision MM R11.1 L3 to resolve the issue. Customers could contact Philips Healthcare Call Center at 1-800-722-9377, #2, #3 and Reference FCO 83000129 for any questions.
Quantity
11 units in US
Distribution
Nationwide Distribution including DC, FL, MI, MO, NM, NY, OH, AND TX
Lot, serial or model codes
Site Numbers:  50310, 84026, 103216, 105518, 505284, 505973, 506122, 533261, 536204, 537203, and 557450.
510(k) numbers
["K051315","K063093"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register