Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541

Recall number
Z-0411-2018
Recalling firm
Baxter Healthcare Corporation
Product
MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541
Status
Terminated
Initiated
2017-12-21
Posted
not on file
Terminated
2020-08-04
Reason
Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.
Root cause
Nonconforming Material/Component
Action
The firm initiated their recall on 12/21/2017 by letter. The recall notice stated the following actions to be taken by the customers: "1. Locate and remove all affected product from your facility. The product code and lot number can be found on the individual MARS Treatment Kit box. 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. Baxter Healthcare Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. Please have your Baxter 8-digit ship-to account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you purchased this product directly from Baxter, complete the enclosed Baxter customer reply form and return it to Baxter by faxing it to 224-270- 5457 or scanning and e-mailing it to fca@baxter.com, even if you do not have any inventory."
Quantity
186 units
Distribution
US, Germany, Sweden, Italy, Belgium, Portugal, Spain, France, Russia, Poland, Turkey, China, Thailand, Taiwan, Hong Kong, Malaysia, Korea, and Mexico.
Lot, serial or model codes
UDI: 37332414107728; Lot Numbers: 0000022597, 0000022648, 0000022663, 0000022671, 0000022721, 0000022767
510(k) numbers
["K133807"]
PMA numbers
not on file
Product code
FLD
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register