Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: Smith & Nephew Taylor Spatial Frame Website software, Catalog Number 71070401. A web-based and mobile application software component of a multilateral external fixation system that is intended for fracture fixation, fixation of long bones and for joint fusions and limb lengthening or deformity corrections which involve cutting of the bone.

Recall number
Z-0409-2018
Recalling firm
Smith & Nephew, Inc.
Product
Smith & Nephew Taylor Spatial Frame Website software, Catalog Number 71070401. A web-based and mobile application software component of a multilateral external fixation system that is intended for fracture fixation, fixation of long bones and for joint fusions and limb lengthening or deformity corrections which involve cutting of the bone.
Status
Terminated
Initiated
2017-12-13
Posted
2018-01-08
Terminated
2020-10-05
Reason
In rare circumstances, when surgeons use the software to plan the strut adjustment, an inverted image of the frame is generated by the software. The frame will appear inverted in the preview images on the Strut Settings tab.
Root cause
Under Investigation by firm
Action
The firm initiated their recall by email on 12/13/2017. Users are asked if they notice an inverted frame in the software to repeat the process for the same case and notify a Smith & Nephew representative. In addition, they are asked to double check plannings for strut adjustments and compare with the images printed as described above.
Quantity
N/A
Distribution
Worldwide
Lot, serial or model codes
Version 5.2.5 and 5.2.6
510(k) numbers
["K142520"]
PMA numbers
not on file
Product code
OSN
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register