Skip to the content

FDA Medical Device Recalls (openFDA)

KCI USA, Inc.: RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic or lumbar fracture, Cervical traction, Skeletal traction

Recall number
Z-0409-2013
Recalling firm
KCI USA, Inc.
Product
RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic or lumbar fracture, Cervical traction, Skeletal traction
Status
Terminated
Initiated
2012-06-27
Posted
2012-11-20
Terminated
2013-04-22
Reason
The firm is initiating a voluntary medical device correction for the RotoRest Delta Therapy System head pack. KCI has received complaints alleging patients experiencing skin breakdown on the head and face due to the positioning of the head pack. In a few instances, the patient received injuries ranging from Stage III to Stage IV pressure ulcers.
Root cause
Component design/selection
Action
KCI sent an Urgent - Voluntary Medical Device Correction letter dated June 27, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that KCI would contact them soon to coorinate a shipment of a replacement of the affected product. For any questions customers were instructed to call 1-800-275-4524, Selection Option 3 followed by Option 2. For questions regarding this recall call 1-800-275-4524.
Quantity
282 units
Distribution
Nationwide Distribution including AL, CA, CO, OK and GA.
Lot, serial or model codes
Model Number 208030, Serial Numbers: 00157087, 00157091, 00157288, 00157375, 00157405, RRCKOO011, RRHKOO065, RRHKOO074, RRHKOO096, RRHK00109, and RRHKOO117.
510(k) numbers
not on file
PMA numbers
not on file
Product code
IOQ
Firm FEI
1625774
Firm city
San Antonio
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register