FDA Medical Device Recalls (openFDA)
KCI USA, Inc.: RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic or lumbar fracture, Cervical traction, Skeletal traction
- Recall number
- Z-0409-2013
- Recalling firm
- KCI USA, Inc.
- Product
- RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic or lumbar fracture, Cervical traction, Skeletal traction
- Status
- Terminated
- Initiated
- 2012-06-27
- Posted
- 2012-11-20
- Terminated
- 2013-04-22
- Reason
- The firm is initiating a voluntary medical device correction for the RotoRest Delta Therapy System head pack. KCI has received complaints alleging patients experiencing skin breakdown on the head and face due to the positioning of the head pack. In a few instances, the patient received injuries ranging from Stage III to Stage IV pressure ulcers.
- Root cause
- Component design/selection
- Action
- KCI sent an Urgent - Voluntary Medical Device Correction letter dated June 27, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that KCI would contact them soon to coorinate a shipment of a replacement of the affected product. For any questions customers were instructed to call 1-800-275-4524, Selection Option 3 followed by Option 2. For questions regarding this recall call 1-800-275-4524.
- Quantity
- 282 units
- Distribution
- Nationwide Distribution including AL, CA, CO, OK and GA.
- Lot, serial or model codes
- Model Number 208030, Serial Numbers: 00157087, 00157091, 00157288, 00157375, 00157405, RRCKOO011, RRHKOO065, RRHKOO074, RRHKOO096, RRHK00109, and RRHKOO117.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- IOQ
- Firm FEI
- 1625774
- Firm city
- San Antonio
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28