Diagnostica Stago, Inc.: STA Liatest Free Protein S, cat. 00516. These kits are intended for use with analyzers of the STA brand name for the quantitative determination of free protein S in plasma by the immuno-turbidimetric method.

Recall number
Z-0408-06
Recalling firm
Diagnostica Stago, Inc.
Product
STA Liatest Free Protein S, cat. 00516. These kits are intended for use with analyzers of the STA brand name for the quantitative determination of free protein S in plasma by the immuno-turbidimetric method.
Status
Terminated
Initiated
2005-10-26
Posted
2006-01-20
Terminated
2006-03-01
Reason
The mean levels (%) of normal and abnormal control plasmas (STA Liatest Control N and P) obtained with STA Liatest Free Protein S are found in the lower part of the range.
Root cause
Other
Action
Notification via recall letter was sent out on 10/26/2005. The letter was unsatisfactory and the letter was resent on 11/10/2005 as a recall letter.
Quantity
150 units
Distribution
Product was distributed to 23 direct account customers nationwide.
Lot, serial or model codes
Lot 0422671 exp. March 2006.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GGP
Firm FEI
2245451
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register