FDA Medical Device Recalls (openFDA)

Vital Signs Colorado Inc.: Vital Signs Disposable General Purpose 9 French Temperature Probe For monitoring of core temperature

Recall number
Z-0407-2013
Recalling firm
Vital Signs Colorado Inc.
Product
Vital Signs Disposable General Purpose 9 French Temperature Probe For monitoring of core temperature
Status
Terminated
Initiated
2012-08-17
Posted
2012-11-20
Terminated
2013-06-14
Reason
There is a potential for epistaxis during use of the Disposable General Purpose 9 French Temperature Probe.
Root cause
Device Design
Action
The firm,GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter and Confirmation form on August 17, 2012 via Fed Ex to its consignees/customers. The letter describes the product, problem and action to be taken. The customers were instructed to do the following: Do not use the Disposable General Purpose 9 French Temperature Probe P/N 1016 or 1016EU; return all unopened product(s) to Vital Signs Devices (VSD) 11039 East Lansing Circle, Englewood, CO 80112-note: replacements and/or refunds discussed in letter are available only in exchange for the return of any unopened product(s); complete and return the URGENT MEDICAL DEVICE CORRECTION CONFIRMATION via fax to: 800-535-7923; forward a copy of the letter to any institutions they may have forwarded the product and for customers using the GXXXXX64 kit that includes the Disposable General Purpose 9 French Temperature Probe follow the directions listed in the letter. Note: VSD is removing p/n 1016 and 1016EU, Disposable General Purpose 9 French Temperature Probe, from the market and offering a replacement. If you have any questions or concerns regarding this notification, contact Customer Service at +1-800-932-0760 (Domestic & International). Hours of Operation 8:00 am EST to 6:00 pm EST.
Quantity
243,685 units
Distribution
Worldwide distribution: USA (nationwide) including states of: AR, CA, FL,GA, IL, LA, MA, MI, MO, NC, NH, NY, OH, OK, PA, RI, SE,TX, and WA; and countries of: AUSTRALIA, BARCELONA, GERMANY, NEW ZEALAND, NETHERLANDS, SWITZERLAND, and UNITED KINGDOM.
Lot, serial or model codes
Model Numbers 1016, 1016EU, R1016ES; GXXXX64 All units Made in Costa Rica are affected Lot Numbers:    12089A 12131A   5575A 5586X 5633U 5664E 5685G 5712T  12090B 12138A 5578K 5603R 5636Y 5666R 5685H  5714Z  12100A 12157B 5578L 5606U 5638P 5668N 5685J  12101B 12173A 5580Y 5614N 5638Q 5669G 5689Y  12102A 5573Q 5580Z 5615N 5648T 5669H 5691T  12108B 5574A 5581A 5621U 5648U 5671D 5692H  12109B 5574B 5581L 5623H 5653Z 5673V 5696Q  12114B 5574X 5583E 5631C 5658D 5683V 5700J  12117B 5574Y 5584C 5632V 5660C 5683W 5703N  12130B 5574Z 5584D 5633T 5664D 5683X 5708R
510(k) numbers
["K822277","K925791","K944949"]
PMA numbers
not on file
Product code
KAK
Firm FEI
1718887
Firm city
Englewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register