Skip to the content

FDA Medical Device Recalls (openFDA)

Intuitive Surgical, Inc.: Da Vinci Single-Site Wristed Needle Driver, REF: 478115, 5 mm

Recall number
Z-0406-2024
Recalling firm
Intuitive Surgical, Inc.
Product
Da Vinci Single-Site Wristed Needle Driver, REF: 478115, 5 mm
Status
Open, Classified
Initiated
2023-10-06
Posted
2023-11-28
Terminated
not on file
Reason
There is the potential for the needle drive to become dislodged and/or broken which could result in uncontrollable movement.
Root cause
Process control
Action
On October 24, 2023, Intuitive issued a "Urgent Medical Device Recall notification" to all affected consignees via E.Mail. Intuitive asked consignees to take the following actions: 1. Read and understand the contents of the letter. 2. Locate and return all affected lots of 478115-03 in your inventory via the standard RMA process by calling Intuitive Customer Service based on your region below. 3. Notify all surgeons, personnel, and sites using da Vinci X and da Vinci Xi Single Site Wristed Needle Driver, PN 478115-03 that they should review and understand the contents of this letter. 4. Complete the attached Acknowledgement Form immediately and return it via email to Intuitive as instructed on the form. a. Please retain a copy of this letter and the acknowledgement form for your files. 5. Inform Intuitive of any Adverse Events* or quality problems concerning the use of the subject device via the standard complaint process. 6. For Customers in the US: 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. a. Complete and submit the report Online Regular Mail or Fax: Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178. 8. Inform all affected personnel at your site who use da Vinci procedures when the return has been completed.
Quantity
317 units
Distribution
US: IL, FL, ID, SC, MI, TX, TN, NY, OR, CO, CA, MS, NJ, Puerto Rico OUS: South Korea
Lot, serial or model codes
UDI:00886874113752/ Serial Numbers: T10220412, T10220420, T10220426, T10220503, T10220518, T10220629, T10220715, T10220721, T10220811, T10220819, T10220901, T10220909, T10220920, T10220923, T10220928, T10221101, T10221109, T10230207, T11221109
510(k) numbers
["K152448"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register