Skip to the content

FDA Medical Device Recalls (openFDA)

Galil Medical, Inc.: Visual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The Visual-ICE System uses high-pressure argon gas that circulates through closed-tip cryoablation needles to induce tissue freezing. Active tissue thawing is achieved by circulating helium gas through the needles or, alternatively , by the use of Galil Medical I-Thaw technology in which a heating element inside the cryoablation needle can be energized to cause thawing.

Recall number
Z-0404-2020
Recalling firm
Galil Medical, Inc.
Product
Visual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The Visual-ICE System uses high-pressure argon gas that circulates through closed-tip cryoablation needles to induce tissue freezing. Active tissue thawing is achieved by circulating helium gas through the needles or, alternatively , by the use of Galil Medical I-Thaw technology in which a heating element inside the cryoablation needle can be energized to cause thawing.
Status
Terminated
Initiated
2019-10-25
Posted
not on file
Terminated
2020-09-16
Reason
Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the needle clogging and/or inadequate freezing performance.
Root cause
Environmental control
Action
An Urgent Medical Device Correction letter was emailed to customers beginning 10/25/2019. A hard copy of this letter was sent by USPS. The letter identified the affected devices, stated the issue, instructed not to use the IceSeed 1.5 or Ice Rod 1.5 PLUS needles until a preventive maintenance activity, to replace the desiccant tubes, has been done by an authorized service engineer.
Quantity
49 systems
Distribution
AK, AL, AR, AZ, CA, FL, GA, HI, KY, MA, MI, MN, NC, ND, NY, OH, PA, SC, TN, UT, VA, Puerto Rico OUS to include: Argentina, Canada, Taiwan, United Kingdom
Lot, serial or model codes
System Serial Number / PM Kit #: VL0031 / PM0506M, VL0036 / PM0546M, VL0067 / PM0515M, VL0069 / PM0565M, VL0075 / PM0532M, VL0079 / PM0539M, VL0082 / PM0491M, VL0085 / PM0486M, VL0086 / PM0499M, VL0088 / PM0489M, VL0090 / PM0513M, VL0092 / PM0548M, VL0093 / PM0544M, VL0094 / PM0520M, VL0100 / PM0516M, VL0102 / PM0529M, VL0105 / PM0560M, VL0112 / PM0542M, VL0115 / PM0525M, VL0136 / PM0561M, VL0148 / PM0514M, VL0153 / PM0531M, VL0171 / PM0564M, VL0175 / PM0559M, VL0181 / PM0512M, VL0186 / PM0533M, VL0191 / PM0492M, VL0220 / PM0545M, VL0230 / PM0528M, VL0238 / PM0500M, VL0249 / PM0498M, VL0271 / PM0497M, VL0275 / PM0543M, VL0280 / PM0549M, VL0283 / PM0488M, VL0285 / PM0566M, VL0287 / PM0562M, VL0291 / PM0524M, VL0300 / PM0550M, VL0301 / PM0541M, VL0303 / PM0530M, VL0304 / PM0563M, VL0308 / PM0523M, VL0309 / PM0507M, VL0310 / PM0490M, VL0311 / PM0555M, VL0312 / PM0526M, VL0321 / PM0552M, VL0362 / PM0527M.
510(k) numbers
["K113860","K123865","K143564"]
PMA numbers
not on file
Product code
GEH
Firm FEI
3004462490
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register