Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc.-Littleton: Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS IMAGING 02 REFURB Model: BI70002000R

Recall number
Z-0404-2019
Recalling firm
Medtronic Navigation, Inc.-Littleton
Product
Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS IMAGING 02 REFURB Model: BI70002000R
Status
Terminated
Initiated
2018-09-17
Posted
not on file
Terminated
2020-05-28
Reason
O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuit
Root cause
Software design
Action
Medtronic Navigation issued Urgent Medical Device Correction letter on 9/27/18, notification letter advising, Medtronic field service personnel will install 4.1.0 software onto the designated consignee systems, update IFU and product labels as well as install a new designed AC power input circuit within the Mobile View Station {MVS). Questions regarding contact Medtronic Technical Support at 1-800-595-9709.
Quantity
641 units
Distribution
Worldwide Distribution - US Nationwide Foreign: AUSTRALIA AUSTRIA BAHRAIN BANGLADESH BELGIUM BRAZIL CANADA CHILE COLOMBIA DENMARK FINLAND FRANCE GERMANY HR ICELAND INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KOREA, REPUBLIC OF LUXEMBOURG MARTINIQUE MEXICO NETHERLANDS NEW ZEALAND POLAND PORTUGAL PUERTO RICO QATAR SAUDI ARABIA SINGAPORE SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED ARAB EMIRATES UNITED KINGDOM
Lot, serial or model codes
Serial Numbers:C0875 through C2012
510(k) numbers
["K173664"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3004785967
Firm city
Littleton
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register