FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: Olympus ENDOEYE HD II Video Telescopes. Models WA50040A ENDOEYE HD II, 10 mm, 0 degree, WA50042A ENDOEYE HD II 10 mm, 30 degree, W7100603 Repair cable 10 mm 0/30 degree. The ENDOEYE video Telescopes are used with other supporting equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.
- Recall number
- Z-0403-2016
- Recalling firm
- Olympus Corporation of the Americas
- Product
- Olympus ENDOEYE HD II Video Telescopes. Models WA50040A ENDOEYE HD II, 10 mm, 0 degree, WA50042A ENDOEYE HD II 10 mm, 30 degree, W7100603 Repair cable 10 mm 0/30 degree. The ENDOEYE video Telescopes are used with other supporting equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.
- Status
- Terminated
- Initiated
- 2015-11-17
- Posted
- 2015-12-10
- Terminated
- 2017-01-24
- Reason
- A damaged temperature sensor at the ENDOEYE tip caused the distal end to become abnormally hot. Excessive heating of the ENDOEYE distal end could result in patient or user injury.
- Root cause
- Unknown/Undetermined by firm
- Action
- Olympus Corporation of Americas sent a letter dated November 25, 2015, to all affected customers to inspect the serial #'s to identify if the affected lots were in inventory and discontinue of the affected lots. Then return all aftected lots for repair and service. Customers were instructed to contact the Customer Care Center at 800-537-5739, Option 2 for instructions on returninng the ENDOEYE for repair and service. The form should be faxed to 484-896-7128. Customers with questions were instructed to call 484-896-5688.
- Quantity
- 182
- Distribution
- US Nationwide Distribution including these states: NE, IL, FL, TX, OH, PA, IN, CA, NY, AL, ND, GA, AK, MI, WA, MS, NJ, IA, AZ, and RI.
- Lot, serial or model codes
- Serial numbers 622483, 622523, 622553, 622566, 622568, 622569, 622573, 622576, 622614, 622615, 622616, 622617, 622618, 622619, 622620, 622632, 622643, 622644, 622645, 622646, 622647, 622654, 622655, 622656, 622657, 622658, 622664, 622666, 622667, 622687, 622688, 622693, 622704, 622705, 622706, 622710, 622711, 622712, 622718, 622719, 622720, 622722, 622723, 622750, 622753, 622764, 622766, 622767, 622769, 622771, 622773, 622775, 622777, 622778, 622785, 622789, 622795, 622797, 622798, 622801, 622806, 623624, 623626, 623629, 623637, 623638, 623639, 623641, 623661, 623669, 623671, 623672, 623673, 623677, 623687, 623688, 623705, 623709, 623713, 623719, 623720, 623722, 623723, 623729, 623730, 623731, 623735, 623736, 623737, 623738, 623739, 623740, 623741, 623742, 623743, 623747, 623749, 623750, 623751, 623752, 623776, 623778, 623784, 623786, 623787, 623788, 623799, 623800, 623801, 623802, 624004, 624005, 624006, 624008, 624011, 624012, 624016, 624017, 624018, 624019, 624024, 624028, 624030, 624041, 624042, 624046, 624054, 624055, 624056, 624057, 624064, 624066, 624070, 624071, 624073, 624076, 624077, 624080, 624082, 624089, 624099, 624100, 624101, 624104, 624107, 624109, 624110, 624112, 624114, 624117, 624119, 624121, 624151, 624158, 624162, 624167, 624170, 624618, 624622, 624643, 624678, 624679, 624682, 624683, 624720, 622539, 622540, 622570, 622572, 623734, 623748, 624739, and 624741.
- 510(k) numbers
- ["K111788"]
- PMA numbers
- not on file
- Product code
- HET
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28