FDA Medical Device Recalls (openFDA)
Advanced Sterilization Products: STERRAD NX, Product Code: 10033 The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
- Recall number
- Z-0403-2014
- Recalling firm
- Advanced Sterilization Products
- Product
- STERRAD NX, Product Code: 10033 The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
- Status
- Terminated
- Initiated
- 2013-11-11
- Posted
- 2013-11-26
- Terminated
- 2014-01-22
- Reason
- Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.
- Root cause
- Labeling design
- Action
- Advanced Sterilization Products sent an Urgent Medical Device Field Safety Notificaiton letter on November 11, 2013, to all affected customers. The letter informed customers of the problems identified and the actions to be taken. Customers were instructed to provide the notice to anyone in their facility that needs to be informed and post this notice where appropriate. Customers were instructed to contact ASP at 1-888-783-7723 for any suspected problems. ASP is collaborating with KARL STORZ Endoskope to determine the sterilization impact of the suction port lumen material. Customers are instructed to contact KARL STORZ Endoskope Technical Support at (800)421-0837, ext 5350 for alternative reprocessing instructions. For questions regarding this recall call 949-453-6400.
- Quantity
- Total 4,381 units
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Please refer to consignee list
- 510(k) numbers
- ["K042116"]
- PMA numbers
- not on file
- Product code
- MLR
- Firm FEI
- 3003662624
- Firm city
- Irvine
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as