Skip to the content

FDA Medical Device Recalls (openFDA)

Kerr Corporation: Freedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470. Freedom Cordless LED Light System is a Class I Medical Device. The device listing number this product is E115952. The intended use of this device is to illuminate oral structures and operating areas.

Recall number
Z-0402-2013
Recalling firm
Kerr Corporation
Product
Freedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470. Freedom Cordless LED Light System is a Class I Medical Device. The device listing number this product is E115952. The intended use of this device is to illuminate oral structures and operating areas.
Status
Terminated
Initiated
2011-11-17
Posted
2012-11-20
Terminated
2012-11-20
Reason
The firm recalled the Freedom Cordless LED Headlight because some units were assembled with an incorrect component in the control board electronics that may cause some Freedom Cordless LED Headlight units to become warm to the touch; however, no skin damage will occur.
Root cause
Labeling mix-ups
Action
Orascoptic sent an "URGENT: MEDICAL DEVICE RECALL" letter dated November 17, 2011. The letter identified the product, problem, and actions to be taken by the customers. An Acknowledgement Form was included for customers to complete and return to the firm. Contact Orascoptic Customer Care at 608-831-2555 for assistance regarding this notice.
Quantity
1330 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Australia, and Europe.
Lot, serial or model codes
Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470.   Freedom Cordless LED Light System has no expiration date. The affected units were produced and distributed since February 2011.
510(k) numbers
not on file
PMA numbers
not on file
Product code
EBA
Firm FEI
3003848022
Firm city
Middleton
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.