Kerr Corporation: BioPlant, a dental bone grafting material. BioPlant has been assigned the product code LYC (Bone Grafting Material, Synthetic) by the FDA and is classified as a Class II Medical Device. BioPlant is a Class II Medical Device. The Device Listing Number is 0007254. BioPlant is a synthetic, multipurpose bone-grafting material intended to be used in dental applications to maintain and restore the bony alveolar ridge in post-extraction tooth-root sockets or after oral surgery, ridge preservation, ridge augmentation, augmentations in the maxillary sinus, repair of periodontal defects and bone voids associated with dental implant placement.

Recall number
Z-0401-2013
Recalling firm
Kerr Corporation
Product
BioPlant, a dental bone grafting material. BioPlant has been assigned the product code LYC (Bone Grafting Material, Synthetic) by the FDA and is classified as a Class II Medical Device. BioPlant is a Class II Medical Device. The Device Listing Number is 0007254. BioPlant is a synthetic, multipurpose bone-grafting material intended to be used in dental applications to maintain and restore the bony alveolar ridge in post-extraction tooth-root sockets or after oral surgery, ridge preservation, ridge augmentation, augmentations in the maxillary sinus, repair of periodontal defects and bone voids associated with dental implant placement.
Status
Terminated
Initiated
2012-10-25
Posted
2012-11-20
Terminated
2013-09-23
Reason
The firm initiated the recall becaue disturbing the BioPlant product soon after it is placed may halt or delay the ossification process.
Root cause
Labeling design
Action
Kerr Corporation sent an Urgent Medical Device Correction letter and acknowledgement form dated October 25, 2012, to all affected customers via USPS First Class. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: "...review the table with product description , part numbers and lot numbers to determine if they have any of the affected product in their inventory and confirm that they were aware of the additional placement instructions necessary to achieve desired ossification results detailed in the paragraph in the letter. Customers were instructed to contact Kerr Customer Care at 18005377123 if they have any questions. Customers were also instructed to complete and fax back the enclosed Acknowledgement Form to 714-288-4609 in order to confirm their receipt of this correction regardless of whether they have any affected product in their inventory. For questions regarding this recall call 714-516-7400.
Quantity
22,410 units
Distribution
Worldwide Distribution - USA (nationwide) Canada, Europe, Australia and ROW consignees .
Lot, serial or model codes
The BioPlant part numbers and lot numbers affected by this recall are as follows:  Product Description Part Numbers Lot Numbers  BioPlant Curved Syringe, 0.125g 216110 All Production  BioPlant Curved Syringe, 0.25g 216112  BioPlant Curved Svrinoe, 0.50<1 216114 Lots Within Shelf-Life  BioPlant Curved Svrinoe, 0.250 216132 20k  BioPlant Intro Pack 216112S
510(k) numbers
not on file
PMA numbers
not on file
Product code
LYC
Firm FEI
2024312
Firm city
Orange
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register