Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories MPG: ARCHITECT STAT Myoglobin Calibrators; list 02K43-01; each kit contains 6 bottles of calibrators ranging from 0 ng/mL to 1200 ng/mL; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064

Recall number
Z-0400-06
Recalling firm
Abbott Laboratories MPG
Product
ARCHITECT STAT Myoglobin Calibrators; list 02K43-01; each kit contains 6 bottles of calibrators ranging from 0 ng/mL to 1200 ng/mL; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064
Status
Terminated
Initiated
2005-10-17
Posted
2006-01-18
Terminated
2006-05-25
Reason
Some of the kits were shipped at the incorrect shipping temperature. They were shipped ambient, but should have been shipped frozen.
Root cause
Other
Action
Abbott telephoned the laboratory end users on 10/17/05, informing them that Abbott discovered that ARCHITECT STAT Myoglobin Calibrators, lot V96128, were shipped to their facilities using the incorrect shipping conditions between August 29th and September 20th 2005. The kits should have been shipped frozen, but were shipped ambient. The error has been corrected. All shipments made after September 21, 2005 have been shipped correctly. Any data generated to date using this kit of calibrators is acceptable, providing that a valid calibration was achieved and product labeling was followed. However, long term use of this kit may result in calibration failures or controls out of range. The accounts were requested to mark the kit(s) of this lot currently in inventory to distinguish them from the replacement kit(s) Abbott will be sending to them. As a precaution, Abbott requested the labs to discontinue use of these kits and destroy them the replacement kits are received.
Quantity
11 kits
Distribution
Texas, Washington, Montana, Pennsylvania and Kentucky, and internationally to Korea
Lot, serial or model codes
list 02K43-01, lot number V96128
510(k) numbers
["K042920"]
PMA numbers
not on file
Product code
DDR
Firm FEI
1415939
Firm city
Abbott Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register