Skip to the content

FDA Medical Device Recalls (openFDA)

Cardiac Science, Inc.: Powerheart AED G3 Automated External Defibrillator Model 9300P Professional series). Cardiac Science AEDs are manufactured by: Manufacturing: Cardiac Science, Inc., 5474 Feltl Road, Minnetonka, MN 55343-7982.

Recall number
Z-0400-05
Recalling firm
Cardiac Science, Inc.
Product
Powerheart AED G3 Automated External Defibrillator Model 9300P Professional series). Cardiac Science AEDs are manufactured by: Manufacturing: Cardiac Science, Inc., 5474 Feltl Road, Minnetonka, MN 55343-7982.
Status
Terminated
Initiated
2004-11-12
Posted
2005-01-15
Terminated
2006-06-25
Reason
AEDs may contain an electronic component that does not meet its full operating specifications. The AEDs with this suspect electronic component may not functionn properly and may interfere with the AED's ability to delivery therapy. In the event that charge noise is presrent it will delay the first defibrillation shock in the 3 shock sequence.
Root cause
Nonconforming Material/Component
Action
Cardiac Science will contact customers via telephone to set up a time for Clinical Educator to visit the customer and conduct a field test on each AED. AEDs that fail the field test will be replaced either on the spot or within 24 hours. AEDs failing the test are either removed by the Clinical Educator or are returned to Cardiac Science upon receiving the new AED. Cardiac Science anticipates testing to be done by the end of December 2004.
Quantity
193
Distribution
Throughout the United States, Canada and Europe.
Lot, serial or model codes
Model no. 9300P serial numbers: 900290 thru 900870
510(k) numbers
["K040637"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
2112020
Firm city
Minnetonka
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register