FDA Medical Device Recalls (openFDA)
Cardiac Science, Inc.: Powerheart AED G3 Automated External Defibrillator Model 9300P Professional series). Cardiac Science AEDs are manufactured by: Manufacturing: Cardiac Science, Inc., 5474 Feltl Road, Minnetonka, MN 55343-7982.
- Recall number
- Z-0400-05
- Recalling firm
- Cardiac Science, Inc.
- Product
- Powerheart AED G3 Automated External Defibrillator Model 9300P Professional series). Cardiac Science AEDs are manufactured by: Manufacturing: Cardiac Science, Inc., 5474 Feltl Road, Minnetonka, MN 55343-7982.
- Status
- Terminated
- Initiated
- 2004-11-12
- Posted
- 2005-01-15
- Terminated
- 2006-06-25
- Reason
- AEDs may contain an electronic component that does not meet its full operating specifications. The AEDs with this suspect electronic component may not functionn properly and may interfere with the AED's ability to delivery therapy. In the event that charge noise is presrent it will delay the first defibrillation shock in the 3 shock sequence.
- Root cause
- Nonconforming Material/Component
- Action
- Cardiac Science will contact customers via telephone to set up a time for Clinical Educator to visit the customer and conduct a field test on each AED. AEDs that fail the field test will be replaced either on the spot or within 24 hours. AEDs failing the test are either removed by the Clinical Educator or are returned to Cardiac Science upon receiving the new AED. Cardiac Science anticipates testing to be done by the end of December 2004.
- Quantity
- 193
- Distribution
- Throughout the United States, Canada and Europe.
- Lot, serial or model codes
- Model no. 9300P serial numbers: 900290 thru 900870
- 510(k) numbers
- ["K040637"]
- PMA numbers
- not on file
- Product code
- MKJ
- Firm FEI
- 2112020
- Firm city
- Minnetonka
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.