FDA Medical Device Recalls (openFDA)

Medtronic Inc., Cardiac Rhythm and Heart Failure: MyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with Medtronic implanted heart devices and a telecommunications connection to transmit stored implanted heart device data to the physician or clinician.

Recall number
Z-0399-2018
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure
Product
MyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with Medtronic implanted heart devices and a telecommunications connection to transmit stored implanted heart device data to the physician or clinician.
Status
Terminated
Initiated
2017-07-14
Posted
not on file
Terminated
2024-06-28
Reason
Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
Root cause
Error in labeling
Action
Customers were notified via letter on 07/14/2017. Medtronic communicated that an updated Clinician Instructions for Use has been released with the following updated information: In patients with more than one Medtronic implanted hear device, device data may not be able to transmit from any of the implanted devices to the physician via the Monitor; The patient could potentially transmit device data unintentionally from one of the implanted heart devices when attempting to transmit from another device implanted in the same patient; and a Medtronic Programmer can be used to obtain data from each implanted heart device. Instructions included to ensure all appropriate personnel are notified of this correction.
Quantity
1,519,984 enrollments total
Distribution
Domestic: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Internationally: Austria, Belgium, Brunei, China, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kazakhstan, Latvia, Lithuania, Malaysia, Malta, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom
Lot, serial or model codes
Model (UPN/GTIN): 24950 (00643169292284, 00643169292345, 00643169292260, 00643169292529, 00643169292208, 00643169292192, 00643169292222, 00643169292239, 00643169292321, 00643169292215, 00643169292246, 00643169409132, 00643169292253, 00643169292277, 00643169292291, 00643169292307, 00643169292314, 00643169292338); 24950D (643169467163, 00643169441484, 00643169441101, 00643169367173, 00643169367227, 00643169367159, 00643169409156, 00643169367296, 00643169367302, 00643169367258, 00643169367180, 00643169367166, 00643169367210, 00643169367289, 00643169409149, 00643169367197, 00643169367142, 00643169367241, 00643169367203); 24950D1 (00643169486744); 24950E (00643169493667, 00643169493681, 00643169493698, 00643169493674, 00643169451988, 00643169452015, 00643169452107, 00643169452039, 00643169452121, 00643169452008,  00643169452145, 00643169452046, 00643169452022, 00643169452053, 00643169452138, 00643169452060, 00643169451995, 00643169452114, 00643169452152, 00643169452077, 00643169452084, 00643169452091); 24950F (00643169545274, 00643169552074); 24950G (00643169663039 00643169624337, 00643169599987, 00643169599994, 00643169624368, 00643169624351, 00643169624344, 00643169552104); 24950GLQ (00643169552289, 00643169552050, 00643169552067); 24950J (00643169840591, 00643169658011, 00643169658066, 00643169657892, 00643169658073, 00643169786196, 00643169658059, 00643169658028, 00643169658004, 00643169657908, 00643169657915, 00643169657946, 00643169657953, 00643169657960, 00643169657977, 00643169657984, 00643169657991, 00643169657922, 00643169657939, 00643169658042, 00643169658080, 00643169658097, 00643169657885, 00643169705746, 00643169658035); 24950JLQ (00643169658127, 00643169658134, 00643169658141, 00643169947115, 00643169705753); 24950K (00643169725270, 00643169725324, 00643169725201, 00643169725355, 00643169725331, 00643169725317, 00643169725287, 00643169725263, 00643169725294, 00643169725218, 00643169725232, 00643169725249, 00643169725256, 00643169725225, 00643169725300, 00643169725348, 00643169725195); 24950KLQ (00643169725362, 00643169725379, 00643169725386)
510(k) numbers
not on file
PMA numbers
["P890003"]
Product code
NVZ
Firm FEI
2182208
Firm city
Mounds View
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register