Medtronic Inc., Cardiac Rhythm and Heart Failure: MyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with Medtronic implanted heart devices and a telecommunications connection to transmit stored implanted heart device data to the physician or clinician.
- Recall number
- Z-0399-2018
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure
- Product
- MyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with Medtronic implanted heart devices and a telecommunications connection to transmit stored implanted heart device data to the physician or clinician.
- Status
- Terminated
- Initiated
- 2017-07-14
- Posted
- not on file
- Terminated
- 2024-06-28
- Reason
- Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
- Root cause
- Error in labeling
- Action
- Customers were notified via letter on 07/14/2017. Medtronic communicated that an updated Clinician Instructions for Use has been released with the following updated information: In patients with more than one Medtronic implanted hear device, device data may not be able to transmit from any of the implanted devices to the physician via the Monitor; The patient could potentially transmit device data unintentionally from one of the implanted heart devices when attempting to transmit from another device implanted in the same patient; and a Medtronic Programmer can be used to obtain data from each implanted heart device. Instructions included to ensure all appropriate personnel are notified of this correction.
- Quantity
- 1,519,984 enrollments total
- Distribution
- Domestic: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Internationally: Austria, Belgium, Brunei, China, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kazakhstan, Latvia, Lithuania, Malaysia, Malta, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom
- Lot, serial or model codes
- Model (UPN/GTIN): 24950 (00643169292284, 00643169292345, 00643169292260, 00643169292529, 00643169292208, 00643169292192, 00643169292222, 00643169292239, 00643169292321, 00643169292215, 00643169292246, 00643169409132, 00643169292253, 00643169292277, 00643169292291, 00643169292307, 00643169292314, 00643169292338); 24950D (643169467163, 00643169441484, 00643169441101, 00643169367173, 00643169367227, 00643169367159, 00643169409156, 00643169367296, 00643169367302, 00643169367258, 00643169367180, 00643169367166, 00643169367210, 00643169367289, 00643169409149, 00643169367197, 00643169367142, 00643169367241, 00643169367203); 24950D1 (00643169486744); 24950E (00643169493667, 00643169493681, 00643169493698, 00643169493674, 00643169451988, 00643169452015, 00643169452107, 00643169452039, 00643169452121, 00643169452008, 00643169452145, 00643169452046, 00643169452022, 00643169452053, 00643169452138, 00643169452060, 00643169451995, 00643169452114, 00643169452152, 00643169452077, 00643169452084, 00643169452091); 24950F (00643169545274, 00643169552074); 24950G (00643169663039 00643169624337, 00643169599987, 00643169599994, 00643169624368, 00643169624351, 00643169624344, 00643169552104); 24950GLQ (00643169552289, 00643169552050, 00643169552067); 24950J (00643169840591, 00643169658011, 00643169658066, 00643169657892, 00643169658073, 00643169786196, 00643169658059, 00643169658028, 00643169658004, 00643169657908, 00643169657915, 00643169657946, 00643169657953, 00643169657960, 00643169657977, 00643169657984, 00643169657991, 00643169657922, 00643169657939, 00643169658042, 00643169658080, 00643169658097, 00643169657885, 00643169705746, 00643169658035); 24950JLQ (00643169658127, 00643169658134, 00643169658141, 00643169947115, 00643169705753); 24950K (00643169725270, 00643169725324, 00643169725201, 00643169725355, 00643169725331, 00643169725317, 00643169725287, 00643169725263, 00643169725294, 00643169725218, 00643169725232, 00643169725249, 00643169725256, 00643169725225, 00643169725300, 00643169725348, 00643169725195); 24950KLQ (00643169725362, 00643169725379, 00643169725386)
- 510(k) numbers
- not on file
- PMA numbers
- ["P890003"]
- Product code
- NVZ
- Firm FEI
- 2182208
- Firm city
- Mounds View
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28