Skip to the content

FDA Medical Device Recalls (openFDA)

Bio-Rad Laboratories, Inc.: VARIANT II Link Hemoglobin A1c Program Reorder Pack, 1600 tests, IVD, For the determination of Hemoglobin A1c in human whole blood. Contents: Elution Buffer A , 1 x 2500 mL Elution Buffer B, 1 x 2100 mL Wash/Diluent Solution, 1 x 2500 mL Cartridge Set CD ROM - 1 Calibrator/Diluent Set Whole Blood Primer, 6 x 1 mL Sample vials, 2 x 100 Instruction Manual Bio-Rad Laboratories, Inc., 400 Alfred nobel Drive, Hercules, CA 94547-1803. Intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography (HLCP).

Recall number
Z-0399-2012
Recalling firm
Bio-Rad Laboratories, Inc.
Product
VARIANT II Link Hemoglobin A1c Program Reorder Pack, 1600 tests, IVD, For the determination of Hemoglobin A1c in human whole blood. Contents: Elution Buffer A , 1 x 2500 mL Elution Buffer B, 1 x 2100 mL Wash/Diluent Solution, 1 x 2500 mL Cartridge Set CD ROM - 1 Calibrator/Diluent Set Whole Blood Primer, 6 x 1 mL Sample vials, 2 x 100 Instruction Manual Bio-Rad Laboratories, Inc., 400 Alfred nobel Drive, Hercules, CA 94547-1803. Intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography (HLCP).
Status
Terminated
Initiated
2011-10-25
Posted
2012-01-11
Terminated
2012-02-02
Reason
It was discovered that the Variant II Turbo Link Calibrator Values card contains incorrect values for IFCC units. The numbers are missing a decimal point.
Root cause
Process control
Action
Bio-Rad Laboratories, Inc. sent a Medical Device Correction Notification letter dated October 25, 2011, via Federal Express to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The correction was initiated by the firm, customers were instructed to contact their regional Bio-Rad office for any questions. For questions regarding this recall call 510-741-6114.
Quantity
63
Distribution
Nationwide Distribution including MI, SC, OH, IL, MN, MD, VA, NY and WI.
Lot, serial or model codes
Catalog number: 270-2716: Lot 70211322; 70211432, 70211440; exp 3/31/2012
510(k) numbers
["K070819"]
PMA numbers
not on file
Product code
KRZ
Firm FEI
2915274
Firm city
Hercules
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register