Trumpf Medical Systems, Inc.: OR table TITAN UT, Surgical Table. Surgical bed with Orthopedic Accessories.

Recall number
Z-0398-2012
Recalling firm
Trumpf Medical Systems, Inc.
Product
OR table TITAN UT, Surgical Table. Surgical bed with Orthopedic Accessories.
Status
Terminated
Initiated
2011-01-21
Posted
2012-01-11
Terminated
2012-08-13
Reason
Product is being recalled due to unintended movement on the Titan Tables.
Root cause
Device Design
Action
Trumpf Medical Systems sent an initial notification to customers dated January 24, 2011. A subsequent "URGENT SAFETY NOTICE-UPDATE" letter dated May 26, 2011 was sent to all affected customers. The letter described the product, problem, and actions to be taken by the customers. A Corrective Action Plan includes a software update to be installed by a Trumpf Technical Service Representative. Contact Technical Service at 888-474-9359 for questions regarding this notice.
Quantity
see line 1.
Distribution
Worldwide Distribution.
Lot, serial or model codes
Material number 1465166, All Serial numbers.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FQO
Firm FEI
3003184737
Firm city
Charleston
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register