Baxter Healthcare Renal Div: HomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, R5C8320, R5C8320R

Recall number
Z-0397-05
Recalling firm
Baxter Healthcare Renal Div
Product
HomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, R5C8320, R5C8320R
Status
Terminated
Initiated
2004-12-23
Posted
2005-01-15
Terminated
2007-12-11
Reason
A software anomaly may allow concentrated solution to be infused into the patient when the HomeChoice unit is used with Physioneal Clear-Flex PD products.
Root cause
Other
Action
Baxter notified the Danish Ministry of Health of the problem with the HomeChoice Pro not prompting the patient when using the Physioneal Clear-Flex PD products on 12/23/04. The patients in Denmark and Sweden will be re-evaluate and re-train each patient as needed using enhanced training materials, which will include the importance of and the rationales for breaking the peel-seal on teh Physioneal ClearFlex prior to use and what to do if mis-infusion occurs. Larger, adhesive, colored stickers will be affixed to all applicable HomeChoice Pro machines to further remind patients of the need to break all peel-seals prior to use. Patients will receive a new Pro-Card when their machine is exchanged, and the exchange will be documented. The HomeChoice Pro software will be updated to correct the software anomaly, validated and installed on the HomeChoice Pro machines.
Quantity
165 units
Distribution
Those HomeChoice units in Denmark and Sweden only, where the Physioneal Clear-Flex products are approved for use.
Lot, serial or model codes
Units with software version 8.7 used with Physioneal Clear-Flex PD products
510(k) numbers
["K012988"]
PMA numbers
not on file
Product code
FKX
Firm FEI
3000210122
Firm city
Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register