FDA Medical Device Recalls (openFDA)
Baxter Healthcare Renal Div: HomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, R5C8320, R5C8320R
- Recall number
- Z-0397-05
- Recalling firm
- Baxter Healthcare Renal Div
- Product
- HomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, R5C8320, R5C8320R
- Status
- Terminated
- Initiated
- 2004-12-23
- Posted
- 2005-01-15
- Terminated
- 2007-12-11
- Reason
- A software anomaly may allow concentrated solution to be infused into the patient when the HomeChoice unit is used with Physioneal Clear-Flex PD products.
- Root cause
- Other
- Action
- Baxter notified the Danish Ministry of Health of the problem with the HomeChoice Pro not prompting the patient when using the Physioneal Clear-Flex PD products on 12/23/04. The patients in Denmark and Sweden will be re-evaluate and re-train each patient as needed using enhanced training materials, which will include the importance of and the rationales for breaking the peel-seal on teh Physioneal ClearFlex prior to use and what to do if mis-infusion occurs. Larger, adhesive, colored stickers will be affixed to all applicable HomeChoice Pro machines to further remind patients of the need to break all peel-seals prior to use. Patients will receive a new Pro-Card when their machine is exchanged, and the exchange will be documented. The HomeChoice Pro software will be updated to correct the software anomaly, validated and installed on the HomeChoice Pro machines.
- Quantity
- 165 units
- Distribution
- Those HomeChoice units in Denmark and Sweden only, where the Physioneal Clear-Flex products are approved for use.
- Lot, serial or model codes
- Units with software version 8.7 used with Physioneal Clear-Flex PD products
- 510(k) numbers
- ["K012988"]
- PMA numbers
- not on file
- Product code
- FKX
- Firm FEI
- 3000210122
- Firm city
- Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28