Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Operations, Inc.: COBAS INTEGRA 800 Tina-Quant Hemoglobin A1cDX Gen.2 Hemoglobin A1cDX test Catalog number 04528123160

Recall number
Z-0396-2017
Recalling firm
Roche Diagnostics Operations, Inc.
Product
COBAS INTEGRA 800 Tina-Quant Hemoglobin A1cDX Gen.2 Hemoglobin A1cDX test Catalog number 04528123160
Status
Terminated
Initiated
2016-10-04
Posted
2016-11-11
Terminated
2017-06-05
Reason
.Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results.
Root cause
Storage
Action
On 10/4/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected Direct Consignees and Distributors via courier. The recall notifications included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Direct Consignee Actions Required " Follow the instructions in the Customer Actions for Prevention/Detection of the Issue section of this Urgent Medical Device Correction (UMDC). " Complete the attached fax form (7184-00-1016) and fax it to 1-844-294-7181. " File this Urgent Medical Device Correction (UMDC) for future reference. Questions Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC. Distributor Actions Required " Ensure proper storage conditions (2-8C) are maintained for the reagents listed in the table above, per their respective method sheet. " If you have not directly shipped any of the products in the table above, you do not need to distribute the enclosed UMDC to your customers. " If you have directly shipped any of the products in the table above, please distribute the enclosed UMDC 16-157 to those customers. " Complete the attached fax form (7185-00-1016) and fax it to 1-844-294-7181. Questions Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this distributor letter or the enclosed UMDC 16-157.
Quantity
4976
Distribution
Nationwide.
Lot, serial or model codes
Not applicable
510(k) numbers
not on file
PMA numbers
not on file
Product code
PDJ
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register