Skip to the content

FDA Medical Device Recalls (openFDA)

Integrated Orbital Implants Inc: Perry-Kolberg Titanium Motility/Support System (Threaded Sleeve and Flat Peg), Model Number 100045 The threaded sleeve and flat peg are two components of the P-K Titanium Motility/Support System. The threaded sleeve is an externally threaded cylinder and an internal drilled hole designed to receive a peg. It is placed in the Bioeye Hydroxyapatite Implant with the use of the P-K Titanium threaded sleeve wrench after a hole has been prepared in the implant by drilling with a series of hypodermic needles of gradually increasing sizes. After the threaded sleeve is placed in the implant, any of the P-K Pegs can be inserted. The affected model which is the subject of this recall comes with the flat peg.

Recall number
Z-0396-2007
Recalling firm
Integrated Orbital Implants Inc
Product
Perry-Kolberg Titanium Motility/Support System (Threaded Sleeve and Flat Peg), Model Number 100045 The threaded sleeve and flat peg are two components of the P-K Titanium Motility/Support System. The threaded sleeve is an externally threaded cylinder and an internal drilled hole designed to receive a peg. It is placed in the Bioeye Hydroxyapatite Implant with the use of the P-K Titanium threaded sleeve wrench after a hole has been prepared in the implant by drilling with a series of hypodermic needles of gradually increasing sizes. After the threaded sleeve is placed in the implant, any of the P-K Pegs can be inserted. The affected model which is the subject of this recall comes with the flat peg.
Status
Terminated
Initiated
2006-12-20
Posted
2007-02-07
Terminated
2007-07-12
Reason
The threads on the threaded sleeve were manufactured with left-handed threads, requiring a modification to the surgical technique for proper insertion. Failure to properly identify the affected components and apply the modified insertion technique will result in an inability to insert the sleeve into the implant.
Root cause
Other
Action
All domestic customers were notified by phone call on 12//20/2006 which was followed up with a fax and the letter sent registered mail. Internationally, customers were notified by fax, email and mail. Customers were instructed to check their inventory of the product by lot number and visual inspection. They were instructed to isolate the affected inventory. They were provided with alternate instructions on how to use the product as it is, if needed. They were provided with a reply card to complete and fax back to the company. They were instructed that when replacement product becomes available, they will be contacted to arrange for the exchange of the affected product.
Quantity
234
Distribution
Worldwide, including USA, United Arab Emirates, Canada, Mexico, Costa Rica, Taiwan & Romania.
Lot, serial or model codes
Lot number: M-50021 Product is not serialized.  Note: Lot code assigned by the contract manufacturing firm.
510(k) numbers
["K974203"]
PMA numbers
not on file
Product code
MQU
Firm FEI
3009688114
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register