FDA Medical Device Recalls (openFDA)
Measurement Specialties Inc: Measurement Specialties Reusable Temperature Probe Autoclavable. Sold as Esophageal/Rectal (Adult), Esophageal/Rectal (Pediatric), and Skin (Adult) Intended for use by a licensed physician to continuously monitor human esophageal, rectal, or skin temperature. They are designed to be used with any patient monitoring system designed to accept 400 Series temperature probes.
- Recall number
- Z-0395-2018
- Recalling firm
- Measurement Specialties Inc
- Product
- Measurement Specialties Reusable Temperature Probe Autoclavable. Sold as Esophageal/Rectal (Adult), Esophageal/Rectal (Pediatric), and Skin (Adult) Intended for use by a licensed physician to continuously monitor human esophageal, rectal, or skin temperature. They are designed to be used with any patient monitoring system designed to accept 400 Series temperature probes.
- Status
- Open, Classified
- Initiated
- 2017-03-13
- Posted
- not on file
- Terminated
- not on file
- Reason
- The reusable temperature probes are provided with instructions for sterilization that have not been proven through a validation study.
- Root cause
- Other
- Action
- Customers were notified via mail or email on March 13, 2017. Additional customers were later identified and notified on June 14, 2017. Instructions included to disregard the sterilization instructions listed in the IFU (A08411F)originally provided with probes and instead utilize the updated IFU 08411G which contains validated sterilization instructions. They were also instructed to forward the recall notification to within their firm and to all organizations if further distributed. For further questions, please call (763) 316-4075.
- Quantity
- 6,076 devices total
- Distribution
- USA (nationwide) distribution to the states of : AL, CA, KY, IL, KS, MN, NY, OH, TX, VA and WI. Distributed internationally to: Australia, Canada, China, England, Germany, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, Singapore, South Africa, South Korea, Taiwan, Vietnam
- Lot, serial or model codes
- Part No. 401AC (Lot No. 14K87246, 15M96863, 13E72645, 12D61386, 13D72302, 16B97593, 15C91286, 13B64377, 14J86547, 12B59416, 13A69334, 15K95972, 15A97196, 15E92377, 15F92378, 13L77336, 12D61023, 14G84704, 14K87499, 14F84193, 13L77335, 16B97594, 15C91419, 13B70167, 15C91448, 15E92375, 15K96046, 16A97197, 12B59529, 16B97798, 15K96125, 16B97595, 15L96777, 12D61829, 13D72644, 15A89998, 13F73449, 12M64134, 15L96745, 15D90215, 15H95035, 13F74403, 13L78489, 13L77336, 12H65961, 14G84878, 12F62179, 14J87014, 12B55182, 14F84444, 14J87081, 13L77336, 13L78489, 14F84400, 12J66632, 16A97380, 12M68130, 15E92376, 14L88737, 14L87904, 12K66633, 13H74994, 13H74995, 12B60056, 12C57285, 14H86266, 16A97592, 14J87164, 12D59168, 14M89765, 12A54328, 15A89854, 14F84634, 14G84624, 15M96829, 15K95896); Part No. 402AC (Lot No. 13E73262, 15E92379, 12E57286, 12L68344, 15E93385, 12H65752, 12H65787, 15F92380, 12J65322, 15F92381, 12K67454, 15A89941, 15A89282, 13H75930, 12F55548, 12K67533, 13H75930, 12J66809, 12B59570, 15E92010, 12A50150, 14E83393); Part No. 409AC (Lot No. 12C56215, 12B56215, 15C90920, 15K28358, 12C60069, 13B70035, 16B97312, 13A69520, 14E83602, 13E73264, 12L68584, 11M57610, 12F59571, 15B90832, 15E92560, 12K63370, 16B97614, 12L54648, 15D92011, 12M63372, 14H85773, 16B97615, 15M97109, 16B97805, 12M69229, 15A89939, 15A90196, 12F60692, 14L88520, 14L88519, 13A70145, 15E92382, 15A90466, 13E73450, 15F92383, 13D72925)
- 510(k) numbers
- ["K962070"]
- PMA numbers
- not on file
- Product code
- FLL
- Firm FEI
- 3012118599
- Firm city
- Andover
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28