Skip to the content

FDA Medical Device Recalls (openFDA)

Measurement Specialties Inc: Measurement Specialties Reusable Temperature Probe Autoclavable. Sold as Esophageal/Rectal (Adult), Esophageal/Rectal (Pediatric), and Skin (Adult) Intended for use by a licensed physician to continuously monitor human esophageal, rectal, or skin temperature. They are designed to be used with any patient monitoring system designed to accept 400 Series temperature probes.

Recall number
Z-0395-2018
Recalling firm
Measurement Specialties Inc
Product
Measurement Specialties Reusable Temperature Probe Autoclavable. Sold as Esophageal/Rectal (Adult), Esophageal/Rectal (Pediatric), and Skin (Adult) Intended for use by a licensed physician to continuously monitor human esophageal, rectal, or skin temperature. They are designed to be used with any patient monitoring system designed to accept 400 Series temperature probes.
Status
Open, Classified
Initiated
2017-03-13
Posted
not on file
Terminated
not on file
Reason
The reusable temperature probes are provided with instructions for sterilization that have not been proven through a validation study.
Root cause
Other
Action
Customers were notified via mail or email on March 13, 2017. Additional customers were later identified and notified on June 14, 2017. Instructions included to disregard the sterilization instructions listed in the IFU (A08411F)originally provided with probes and instead utilize the updated IFU 08411G which contains validated sterilization instructions. They were also instructed to forward the recall notification to within their firm and to all organizations if further distributed. For further questions, please call (763) 316-4075.
Quantity
6,076 devices total
Distribution
USA (nationwide) distribution to the states of : AL, CA, KY, IL, KS, MN, NY, OH, TX, VA and WI. Distributed internationally to: Australia, Canada, China, England, Germany, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, Singapore, South Africa, South Korea, Taiwan, Vietnam
Lot, serial or model codes
Part No. 401AC (Lot No. 14K87246, 15M96863, 13E72645, 12D61386, 13D72302, 16B97593, 15C91286, 13B64377, 14J86547, 12B59416, 13A69334, 15K95972, 15A97196, 15E92377, 15F92378, 13L77336, 12D61023, 14G84704, 14K87499, 14F84193, 13L77335, 16B97594, 15C91419, 13B70167, 15C91448, 15E92375, 15K96046, 16A97197, 12B59529, 16B97798, 15K96125, 16B97595, 15L96777, 12D61829, 13D72644, 15A89998, 13F73449, 12M64134, 15L96745, 15D90215, 15H95035, 13F74403, 13L78489, 13L77336, 12H65961, 14G84878, 12F62179, 14J87014, 12B55182, 14F84444, 14J87081, 13L77336, 13L78489, 14F84400, 12J66632, 16A97380, 12M68130, 15E92376, 14L88737, 14L87904, 12K66633, 13H74994, 13H74995, 12B60056, 12C57285, 14H86266, 16A97592, 14J87164, 12D59168, 14M89765, 12A54328, 15A89854, 14F84634, 14G84624, 15M96829, 15K95896);    Part No. 402AC (Lot No. 13E73262, 15E92379, 12E57286, 12L68344, 15E93385, 12H65752, 12H65787, 15F92380, 12J65322, 15F92381, 12K67454, 15A89941, 15A89282, 13H75930, 12F55548, 12K67533, 13H75930, 12J66809, 12B59570, 15E92010, 12A50150, 14E83393);    Part No. 409AC (Lot No. 12C56215, 12B56215, 15C90920, 15K28358, 12C60069, 13B70035, 16B97312, 13A69520, 14E83602, 13E73264, 12L68584, 11M57610, 12F59571, 15B90832, 15E92560, 12K63370, 16B97614, 12L54648, 15D92011, 12M63372, 14H85773, 16B97615, 15M97109, 16B97805, 12M69229, 15A89939, 15A90196, 12F60692, 14L88520, 14L88519, 13A70145, 15E92382, 15A90466, 13E73450, 15F92383, 13D72925)
510(k) numbers
["K962070"]
PMA numbers
not on file
Product code
FLL
Firm FEI
3012118599
Firm city
Andover
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register