Baxter Healthcare Renal Div: Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA
- Recall number
- Z-0395-05
- Recalling firm
- Baxter Healthcare Renal Div
- Product
- Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA
- Status
- Terminated
- Initiated
- 2003-05-28
- Posted
- 2004-12-02
- Terminated
- 2008-02-26
- Reason
- The air detector may not detect air bubbles consistently at the selected limit.
- Root cause
- Other
- Action
- Important Product Information letters were sent to customers on 5/28/03, informing them of the potential for the air detector not detecting air bubbles consistently. The letters instructed the accounts to measure the air detector sensor voltage output on each of their instruments, using the enclosed instructions, to determine whether the installation of an air detector attenuator connector is needed. The instructions also included directions for installation of the attenuator connector. Since the instrument requires re-qualification after the testing and installation, the accounts were given the choice to have the testing and installation done during the next service call or as part of the instrument''s next normally scheduled maintenance.
- Quantity
- 28,482 units
- Distribution
- The devices were distributed nationwide and internationally to Baxter subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Denmark, Ecuador, Finland, France, the Republic of Georgia, Greece, Guatemala, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Korea, Lebanon, Malaysia, Morocco, the Netherlands, Oman, Palestine, Panama, Peru, the Philippines, Poland, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Tunisia, Turkey and the United Kingdom.
- Lot, serial or model codes
- All System 1000 units with serial numbers prior to 22224.
- 510(k) numbers
- ["K910215"]
- PMA numbers
- not on file
- Product code
- FKP
- Firm FEI
- 3000210122
- Firm city
- Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28