FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: Recovery G2 Filter System - Jugular/Subclavian Delivery Kit, for use in the prevention of recurrent pulmonary embolism. Catalog Number RF-220J ***For Canada & Australia only***

Recall number
Z-0394-2007
Recalling firm
Bard Peripheral Vascular Inc
Product
Recovery G2 Filter System - Jugular/Subclavian Delivery Kit, for use in the prevention of recurrent pulmonary embolism. Catalog Number RF-220J ***For Canada & Australia only***
Status
Terminated
Initiated
2006-12-20
Posted
2007-02-01
Terminated
2008-10-23
Reason
This action is being taken because BPV has recently received three complaints for the device that involve reports of introducer sheath tip damage. The damage may result in loss of tip integrity.
Root cause
Other
Action
Recall letters were sent to all consignees via Federal Express overnight delivery with signature receipt required on December 20, 2006. Each letter included a Recall and Effectiveness Check Form for each consignee to fill out and fax to Bard Peripheral Vascular, Inc (BPV). Customers will be asked to immediately remove product from inventory and isolated from use and contact BARD for return authorization. All returned product will be destroyed.
Quantity
124 units
Distribution
Worldwide including USA, Australia, and Canada
Lot, serial or model codes
Lot numbers: GFPK3279, GFPK3280, GFPK3281 & GFPK3282
510(k) numbers
["K052578"]
PMA numbers
not on file
Product code
DTK
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register