Skip to the content

FDA Medical Device Recalls (openFDA)

Reshape Medical Inc: ReShape Integrated Dual Balloon System, Gen 1, Model No. 01-0013-001, Catalog No. RSM110 product Usage: The ReShape Integrated Dual Balloon System is indicated for weight reduction when used in conjunction with diet and exercise, in obese patients with a Body Mass Index (BMI) of 30 40 kg/m2 and one or more obesityrelated comorbid conditions. It is indicated for use in adult patients who have failed weight reduction with diet and exercise alone.

Recall number
Z-0392-2017
Recalling firm
Reshape Medical Inc
Product
ReShape Integrated Dual Balloon System, Gen 1, Model No. 01-0013-001, Catalog No. RSM110 product Usage: The ReShape Integrated Dual Balloon System is indicated for weight reduction when used in conjunction with diet and exercise, in obese patients with a Body Mass Index (BMI) of 30 40 kg/m2 and one or more obesityrelated comorbid conditions. It is indicated for use in adult patients who have failed weight reduction with diet and exercise alone.
Status
Terminated
Initiated
2016-10-14
Posted
2016-11-10
Terminated
2017-01-24
Reason
ReShape Medical is recalling the Integrated Dual Balloon System Gen 1 due to the potential leak of distillation fluid during balloon inflation.
Root cause
Device Design
Action
ReShape sent a customer notification letter dated October 14, 2016 was sent to customers to inform them that ReShape Medical is recalling the Integrated Dual Balloon System due to customer reports of instillation fluid leaking during the device inflation procedure. The letter informs the customers that any instillation fluid leakage during the inflation process will cause an incomplete balloon fill and require immediate removal of the device. Customers are instructed to contact ReShape Medical Customer Service at (844) 9377374 for any questions related to the product returns.
Quantity
114 units
Distribution
US Nationwide Distribution in the states of: MO, TX, AZ, CA, NY, TN, GA, LA, FL, MA, OH, IN, IL
Lot, serial or model codes
Lot Number Expiration Date 160815-001 2018-08-07 160816-001 2018-08-07 160824-003 2018-08-18 160825-001 2018-08-18 160825-002 2018-08-18
510(k) numbers
not on file
PMA numbers
["P140012S002"]
Product code
LTI
Firm FEI
3013508647
Firm city
San Clemente
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register