Baxter Healthcare Corp.: Baxter, VIAL-MATE Adaptor, 2B8071, Rx Only, STERILE R, Baxter Healthcare Corporation, Deerfield, IL 30015 USA The VIAL-MATE is reconstitution device labeled for use with 20 mm drug vials and 50mL, 100mL, and 250mL Single Pack Mini Bags.
- Recall number
- Z-0392-2014
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Baxter, VIAL-MATE Adaptor, 2B8071, Rx Only, STERILE R, Baxter Healthcare Corporation, Deerfield, IL 30015 USA The VIAL-MATE is reconstitution device labeled for use with 20 mm drug vials and 50mL, 100mL, and 250mL Single Pack Mini Bags.
- Status
- Terminated
- Initiated
- 2013-09-25
- Posted
- 2013-11-26
- Terminated
- 2014-07-22
- Reason
- Incomplete foil seal on one lot of sterile product.
- Root cause
- Packaging process control
- Action
- Baxter Healthcare sent an Urgent Product Recall letter dated September 25, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate and remove all affected product from their facility, contact Baxter's Center for Service at 1-888-229-0001 to arrange for return and credit. Complete the attached Customer Reply Form and return it to Baxter by fax or email. Customers with questions were instructed to call 1-800-422-9837. For questions regarding this recall call 224-948-4770.
- Quantity
- 38,400 units
- Distribution
- Nationwide Distribution including WV, KY, IA, AR, OH, IL, IN, MI, KS, TN, WI, MN, MO.
- Lot, serial or model codes
- Lot Number GR13G23028
- 510(k) numbers
- ["K973654"]
- PMA numbers
- not on file
- Product code
- LHI
- Firm FEI
- 1417572
- Firm city
- Deerfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28