Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Gastroenterology / Urology: Medtronic Slimline Single-Use Internal Reference pH Catheter for Gastroesophageal pH measurements. Models 9012P3101, 9012P3111, 9012P3121, 9012P3131 and 9012P3141. Distributed by / Medtronic Inc. 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604 USA

Recall number
Z-0392-06
Recalling firm
Medtronic Gastroenterology / Urology
Product
Medtronic Slimline Single-Use Internal Reference pH Catheter for Gastroesophageal pH measurements. Models 9012P3101, 9012P3111, 9012P3121, 9012P3131 and 9012P3141. Distributed by / Medtronic Inc. 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604 USA
Status
Terminated
Initiated
2005-12-09
Posted
2006-01-13
Terminated
2006-12-21
Reason
Certain lots have a manufacturing defect that allows the water pushed through the catheter to travel back into the connector of the Digitrapper pH recorder. Medtronic has determined that patient safety is not compromised, but catheter function may be affected and damage may result to the Digitrapper pH recorder.
Root cause
Other
Action
Letters will be sent to distributors, end-users and Medtronic Field Personnel. The end-user will be asked to return via fax a confirmation/customer receipt. Letters identify the affected catalog number and lot numbers. Customers are asked to return affected product and will receive replacement product.
Quantity
140 catheters
Distribution
CA, CO, FL, MA, OH, PA. OUS to include Finland, Italy and Portugal
Lot, serial or model codes
Lot numbers 571, 606, 695, 754, 952, 945, 1024, 1066
510(k) numbers
not on file
PMA numbers
not on file
Product code
FFT
Firm FEI
1000135098
Firm city
Shoreview
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register