FDA Medical Device Recalls (openFDA)
Maquet Medical Systems USA: Accessory Adapter Part number 600525A0 Product Usage: designed for the mounting of MAQUET accessories with the dovetail guide interface immediately before, during and after surgical interventions as well as for examination and treatment.
- Recall number
- Z-0390-2017
- Recalling firm
- Maquet Medical Systems USA
- Product
- Accessory Adapter Part number 600525A0 Product Usage: designed for the mounting of MAQUET accessories with the dovetail guide interface immediately before, during and after surgical interventions as well as for examination and treatment.
- Status
- Terminated
- Initiated
- 2016-10-05
- Posted
- not on file
- Terminated
- 2018-02-02
- Reason
- Maquet Inc. is initiating a voluntary field action on the Accessory Adapter part number 600525A0 due to a complaint in which a head rest mounted to the accessory adapter slipped out of the mountain rail.
- Root cause
- Nonconforming Material/Component
- Action
- Maquet Medical Systems USA sent an Urgent Product Recall Medical Device Removal letter dated October 5, 2016 to the affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: 1. Read this Medical Device Removal Notice and disseminate the information provided to all the users of the device. 2. Examine your inventory immediately to determine if you have the Accessory Adapter(s). If so, remove the affected products, quarantine them and keep them in a secure location for shipment back to Maquet. 3. Please fill out and sign the enclosed Medical Device Removal Response Form on page 4 and return it either via email to FieldActions@maquet.com or fax to 1-973-909-9927. For question contact your local MAQUET Representative or call MAQUET Customer Service at 1-888-627-8383 (Press option 2, then followed by option 3) Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. EDT.
- Quantity
- 9 Units (US) 141 Units (OUS)
- Distribution
- Worldwide Distribution - US Nationwide in the states of AZ, CA, DC, MI, MO, and PA
- Lot, serial or model codes
- 600525A0
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FQO
- Firm FEI
- 3008355164
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28