Skip to the content

FDA Medical Device Recalls (openFDA)

Xstrahl Limited: Treatment applicator for Gulmay Medical 150 or Xstrahl 150 superficial X-ray therapy system

Recall number
Z-0389-2022
Recalling firm
Xstrahl Limited
Product
Treatment applicator for Gulmay Medical 150 or Xstrahl 150 superficial X-ray therapy system
Status
Terminated
Initiated
2021-08-27
Posted
not on file
Terminated
2026-03-04
Reason
Base plate may detach from the main body of the treatment applicator.
Root cause
Device Design
Action
The firm notified their consignees by email on 08/27/2021. The notice requested the following actions be taken: Xstrahl Limited are issuing the following inspection advice for all X150 specifical system users. 1. Prior to any clinical exposure the operator of the system should inspect the treatment applicator prior to inserting it into the treatment head of the X150. 2. As part of the daily assurance process the treatment applicators should be inspected. 3. If a treatment applicator is dropped it should be inspected by medical physics to ensure it is not damaged. 4. If the base plate is missing the applicators must be withdrawn from clinical use until it is repaired and accepted by medical physics. 5. In the case of an applicator where the base plate has become detached, review the prescription history for the patients for whom that applicator was prescribed. 6. Please advise Xstrahl Limited if you believe patients have been treated with a base plate missing, with detail of the applicator size and clinical filter, kV and HVL, and we will contact you directly.
Quantity
120 applicators
Distribution
US and ALGERIA, AUSTRALIA, AUSTRIA, BULGARIA, CANADA, COSTA RICA, CROATIA, CYPRUS, DENMARK, FRANCE, GERMANY, HONG KONG, IRELAND, ITALY, KAZAKHSTAN, LATVIA, MALTA, MEXICO, NEW ZEALAND, NORWAY, PANAMA, ROMANIA, RUSSIA, SAUDI ARABIA, SPAIN, SWEDEN, SWITZERLAND, UK, UKRAINE
Lot, serial or model codes
GM0072, GM0077, GM0086, GM0117, GM0131, GM0158, GM0184, GM0189, GM0275, GM0307, GM0338, GM0379, GM0414, GM0469, GM0506
510(k) numbers
["K962613"]
PMA numbers
not on file
Product code
JAD
Firm FEI
3004561814
Firm city
Walsall
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register