Smith & Nephew Inc: Oxinium Femoral Head, 36 MM OD-3, 12/14 Taper, REF 71343603, Sterile, Smith & Nephew, Memphis TN 38116
- Recall number
- Z-0389-2010
- Recalling firm
- Smith & Nephew Inc
- Product
- Oxinium Femoral Head, 36 MM OD-3, 12/14 Taper, REF 71343603, Sterile, Smith & Nephew, Memphis TN 38116
- Status
- Terminated
- Initiated
- 2009-01-23
- Posted
- 2009-11-20
- Terminated
- 2012-02-07
- Reason
- Radius at the end of the taper of the femoral implant is smaller than specification. This error could result in the tapers not locking fully or properly on femoral stems or in causing impingement and/or increased wear debris. The deficiency may not be noticed by the user. Impingement and/or wear debris may lead to revision.
- Root cause
- Other
- Action
- All affected Smith & Nephew Sales Representatives were notified of problem and the recall overnight mail on 01/23/2009. All affected Smith & Nephew International Distributors were notified via email and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return. Further questions are directed to the contacts at Smith and Nephew listed in the letter.
- Quantity
- 925 units
- Distribution
- Nationwide; 17 international: Switzerland, Dubai, Brazil, Canada, South Africa, Australia, Belgium, Spain, Thailand Turkey, Costa Rica, Denmark, India, Italy, and France.
- Lot, serial or model codes
- Lot Numbers: 08BM09885, 08EM06396, 08FM14910, 08HM10480, 08KM07754, 08KM12526, 08HM03206, 08GM03007R, 08EM19444, 08EM19445, 08EM19446, 08EM20186, 08FM00231, 08RM00242, 08FM03526, 08FM04628, 08FM05689, 08FM06247, 08FM14906, 08FM17347, 08FM22085, 08FM22087, 08GM03007, 08GM03489, 08GM06660, 08GM11382, 08GM11425, 08GM12094, 08GM13061, 08GM17805, 08GM18609, 08HM02631, 08HM03926, 08HM10037, 08HM13107, 08HM13118, 08HM16147, 08HM16150, 08HM17621, 08HM18516, 08JM00370, 08JM04970, 08JM04971, 08JM05505, 08JM05506, 08JM08122, 08JM09041, 08JM11262, 08JM14571, 08JM15417, 08JM19246, 08JM19247, 08JM19248, 08JM20328, 08JM21716, 08JM21717, 08JM21718, 08KM03713, 08KM07732, 08KM14807, 08KM20680, 08KM23716, 08LM01328, 08LM02300, 08LM11759, 08LM12842, and 08LM13924
- 510(k) numbers
- ["K030766"]
- PMA numbers
- not on file
- Product code
- JDI
- Firm FEI
- 1020279
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28