FDA Medical Device Recalls (openFDA)
Hospira Inc: Hospira brand Plum A+I.V. Infusion Pump, List No. 11971-04; Hospira brand Plum A+I.V. Infusion Pump, List No. 12391-04; Hospira brand Plum A+3 I.V. Infusion Pump, List No. 12618-04; Hospira brand Plum A+3 I.V. Infusion Pump, List No. 12348-04; Infusion pump is a multi-function fluid delivery infusion system, consisting of pumping module and assortment of IV sets. Product is manufactured and distributed by Hospira Inc., 755 Jarvis Dr., Morgan Hill, CA
- Recall number
- Z-0389-06
- Recalling firm
- Hospira Inc
- Product
- Hospira brand Plum A+I.V. Infusion Pump, List No. 11971-04; Hospira brand Plum A+I.V. Infusion Pump, List No. 12391-04; Hospira brand Plum A+3 I.V. Infusion Pump, List No. 12618-04; Hospira brand Plum A+3 I.V. Infusion Pump, List No. 12348-04; Infusion pump is a multi-function fluid delivery infusion system, consisting of pumping module and assortment of IV sets. Product is manufactured and distributed by Hospira Inc., 755 Jarvis Dr., Morgan Hill, CA
- Status
- Terminated
- Initiated
- 2005-05-18
- Posted
- 2006-01-12
- Terminated
- 2006-12-21
- Reason
- The pump may shut-off prematurely during battery operation without providing adequate response time when 'E320' service warning message appears.
- Root cause
- Other
- Action
- On 5/18/05, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected products and providing instructions on the recall.
- Quantity
- not on file
- Distribution
- Product was distributed nationwide to 746 domestic consignees. Product was distributed internationally to the following countries: Puerto Rico, Canada, Saudi Arabia, Taiwan, Australia, Italy, Jordan, New Zealand, Spain, Turkey, and United Arab Emirates.
- Lot, serial or model codes
- List No. 11971-04 Plum A+ I.V. Infusion Pump; List No. 12391-04 Plum A+ I.V. Infusion Pump; List No. 12618-04 Plum A+3 I.V. Infusion Pump; List No. 12348-04 Plum A+3 I.V. Infusion Pump. List Numbers: 12618-04; List Numbers: 11971-36; (shipped 12391-04; (shipped 11971-04; domestically) 11971-04; internationally) 12391-04; 12348-04; 12348-04; 12391-13; 12391-1312391-1 12618-04. All serial numbers of the above list numbered product are included in this correction.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 2921482
- Firm city
- Morgan Hill
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28