Hospira Inc: Hospira brand Plum A+I.V. Infusion Pump, List No. 11971-04; Hospira brand Plum A+I.V. Infusion Pump, List No. 12391-04; Hospira brand Plum A+3 I.V. Infusion Pump, List No. 12618-04; Hospira brand Plum A+3 I.V. Infusion Pump, List No. 12348-04; Infusion pump is a multi-function fluid delivery infusion system, consisting of pumping module and assortment of IV sets. Product is manufactured and distributed by Hospira Inc., 755 Jarvis Dr., Morgan Hill, CA

Recall number
Z-0389-06
Recalling firm
Hospira Inc
Product
Hospira brand Plum A+I.V. Infusion Pump, List No. 11971-04; Hospira brand Plum A+I.V. Infusion Pump, List No. 12391-04; Hospira brand Plum A+3 I.V. Infusion Pump, List No. 12618-04; Hospira brand Plum A+3 I.V. Infusion Pump, List No. 12348-04; Infusion pump is a multi-function fluid delivery infusion system, consisting of pumping module and assortment of IV sets. Product is manufactured and distributed by Hospira Inc., 755 Jarvis Dr., Morgan Hill, CA
Status
Terminated
Initiated
2005-05-18
Posted
2006-01-12
Terminated
2006-12-21
Reason
The pump may shut-off prematurely during battery operation without providing adequate response time when 'E320' service warning message appears.
Root cause
Other
Action
On 5/18/05, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected products and providing instructions on the recall.
Quantity
not on file
Distribution
Product was distributed nationwide to 746 domestic consignees. Product was distributed internationally to the following countries: Puerto Rico, Canada, Saudi Arabia, Taiwan, Australia, Italy, Jordan, New Zealand, Spain, Turkey, and United Arab Emirates.
Lot, serial or model codes
List No. 11971-04 Plum A+ I.V. Infusion Pump; List No. 12391-04 Plum A+ I.V. Infusion Pump;  List No. 12618-04 Plum A+3 I.V. Infusion Pump;  List No. 12348-04 Plum A+3 I.V. Infusion Pump.  List Numbers: 12618-04; List Numbers: 11971-36; (shipped 12391-04; (shipped 11971-04; domestically) 11971-04; internationally) 12391-04;  12348-04; 12348-04;  12391-13; 12391-1312391-1   12618-04. All serial numbers of the above list numbered product are included in this correction.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FRN
Firm FEI
2921482
Firm city
Morgan Hill
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register