Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare Biosciences: DeltaVision OMX SR Imaging System Model Number: 29115476 The DeltaVision OMX SR Imaging System is a piece of laboratory equipment for super resolution imaging offluorescently-la beling biological samples. Techniques such as 3D Structured Illumination, Total Internal Reflection Fluorescence and localization microscopy may be used to examine a variety of samples over long periods of time.

Recall number
Z-0388-2019
Recalling firm
GE Healthcare Biosciences
Product
DeltaVision OMX SR Imaging System Model Number: 29115476 The DeltaVision OMX SR Imaging System is a piece of laboratory equipment for super resolution imaging offluorescently-la beling biological samples. Techniques such as 3D Structured Illumination, Total Internal Reflection Fluorescence and localization microscopy may be used to examine a variety of samples over long periods of time.
Status
Terminated
Initiated
2018-09-10
Posted
not on file
Terminated
2020-05-12
Reason
Non-conformance with the DeltaVision OMX SR, the interlocks are not wired in a redundant fashion to make the interlock system single-fault tolerant.
Root cause
Device Design
Action
The firm, GE Healthcare, sent an "Important Electronic Product Radiation Warning" letter to affected customers on September 10, 2018. The letter identified the affected product, problem and actions to be taken. GE Healthcare will, without charge, remedy the defect or bring the product into compliance. Your GE Service Representative will contact you to implement the correction. For questions contact your local GEHC service representative or call 425-657-1512 or email: trisha.koenke@ge.com.
Quantity
39
Distribution
Worldwide - US Nationwide Distribution
Lot, serial or model codes
System Serial number OM20072 - OM40122
510(k) numbers
not on file
PMA numbers
not on file
Product code
REM
Firm FEI
1281317
Firm city
Westborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register