Philips Medical Systems (Cleveland) Inc: Brilliance iCT Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Recall number
Z-0388-2018
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance iCT Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Status
Terminated
Initiated
2017-11-10
Posted
not on file
Terminated
2020-11-05
Reason
Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.
Root cause
Under Investigation by firm
Action
An Urgent - Field Safety Notice Medical Device Correction letter, dated November 1, 2017, was sent to customers. The letter identified the affected device, described the issue, and discussed the hazard involved. Users are to cease use immediately and contact their Philips Field Service Engineer, if any unusual noises are heard within the gantry, or if the patient support tabletop is in free float. Philips Field Service Engineers will correct the system by replacing the affected fasteners. If further info or support concerning this issue is needed, customers can contact their local Philips rep or the Customer Care Solutions Center at 1-800-722-9377.
Quantity
13
Distribution
Worldwide Distribution -- US, Argentina, China, Republic of Korea, Russian Federation, Slovakia, Turkey, and United Kingdom. 4 in US, (1 military), 14 foreign accounts
Lot, serial or model codes
Code 728306, SN# 100776, UDI: (01)00884838059474(21)100776; 100778, UDI: (01)00884838059474(21)100778; 100779, UDI: (01)00884838059474(21)100779; 100780, UDI: (01)00884838059474(21)100780; 100781, UDI: (01)00884838059474(21)100781; 100782, UDI: (01)00884838059474(21)100782; 100783, UDI: (01)00884838059474(21)100783; 100784, UDI: (01)00884838059474(21)100784; 100785, UDI: (01)00884838059474(21)100785; 100787, UDI: (01)00884838059474(21)100787; 100790, UDI: (01)00884838059474(21)100790; 100791, UDI: (01)00884838059474(21)100791; and  100794, UDI: (01)00884838059474(21)1007.
510(k) numbers
["K162838"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register