Boston Scientific: Boston Scientific Mach 1 Peripheral Guide Catheter, 6F, 55cm, LIMA Tip Shape, UPN # M001196620, Catalog # 19-662. Made in Mexico: Edificio G1 Local B, H, Parque Industrial La Mesa, Tijuana, Baja California 22650 MEX.

Recall number
Z-0388-2007
Recalling firm
Boston Scientific
Product
Boston Scientific Mach 1 Peripheral Guide Catheter, 6F, 55cm, LIMA Tip Shape, UPN # M001196620, Catalog # 19-662. Made in Mexico: Edificio G1 Local B, H, Parque Industrial La Mesa, Tijuana, Baja California 22650 MEX.
Status
Terminated
Initiated
2006-12-13
Posted
2007-02-07
Terminated
2007-06-28
Reason
Excess strands of resin may exist in the lumen near the hub of certain batches/lots of the Mach 1 Guide Catheters. If this excess resin is present in the catheter, there is the potential of an embolization resulting from the excess resin detaching from the device during the procedure.
Root cause
Other
Action
An Urgent Medical Device Recall letter, dated 12/13/06, and a Clarification of Recall Notification letter, dated 12/14/06, were sent to all affected hospitals. The letter describes the issue, affected product, asks for customers to discontinue further distribution or use of any remaining product and asks for product to be returned. Replacement product will be available for any/all returned product.
Quantity
18
Distribution
Nationwide.
Lot, serial or model codes
Batch / Lot #: 687877, 687878, 687879, 688002
510(k) numbers
["K010874"]
PMA numbers
not on file
Product code
DQY
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register