Skip to the content

FDA Medical Device Recalls (openFDA)

Medical Compression Systems: ActiveCare DVT Medical Compression Systems OR Aqiva, IL Intended for use in: Preventing deep Vein Thrombosis; Enhancing blood circulation; Diminishing post-operative pain and swelling; Reducing wound healing time; treatment and assistance in healing of stasis dermatitis, venous stasis ulcers and diabetic leg ulcers; Treatment of chronic venous insufficiency and; Reducing edema

Recall number
Z-0387-2012
Recalling firm
Medical Compression Systems
Product
ActiveCare DVT Medical Compression Systems OR Aqiva, IL Intended for use in: Preventing deep Vein Thrombosis; Enhancing blood circulation; Diminishing post-operative pain and swelling; Reducing wound healing time; treatment and assistance in healing of stasis dermatitis, venous stasis ulcers and diabetic leg ulcers; Treatment of chronic venous insufficiency and; Reducing edema
Status
Terminated
Initiated
2011-08-28
Posted
2012-01-11
Terminated
2012-07-11
Reason
This voluntary field action is being undertaken because MCS has received a limited number of reports involving the use of a non-MCS AC/DC adaptor with the ActiveCare DVT and ActiveCare+S.F.T. devices. The firm found that an ActiveCare DVT device may work improperly if connected to a non-MCS AC/DC adaptor with a higher nominal voltage. As specified in each product's User Manual, the ActiveCare DVT
Root cause
Device Design
Action
Medical Compression Systems (DBN) Ltd, sent out "VOLUNTARY FIELD ACTION" letters dated September 15, 2011 to all affected customers. The letter included description of the device and problem. Customers were instructed to check their inventory and make a list of serial numbers of all applicable ActiveCare+S.F.T. and ActiveCare DVT devices in their inventory, include the list on the attached excel file (Field Action Verification Form) and email to maya@mcsmed.com; even if they do not have any such devices. The letter also states that within 15 days from the email, they will receive instructions for the next steps. In addition, the letter stated -YOU MAY CONTINUE TO USE THE ActiveCare+S.F.T. DEVICES WHILE AWAITINGTHE INSTALLATION OF THE OVP. In case you have questions or you need any assistance, you can contact the firm at (317) 485-7887 or e-mail to: maya@mcsmed.com or mahlet@mcsmed.com.
Quantity
2217 units
Distribution
Nationwide distribution.
Lot, serial or model codes
Device Catalog Number A502B-0001-02 and H502B-0001-07
510(k) numbers
["K023573"]
PMA numbers
not on file
Product code
JOW
Firm FEI
3003715046
Firm city
OR Aqiva
Firm state
not on file
Firm country
Israel

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register