FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corp.: Sabraset Administration Sets with AutoClamp; for use with model 6060 multi-therapy pumps only; a prescription, sterile, non-pyrogenic fluid pathway; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., the following sets were made in Mexico: a) Sabraset 56050 100 Administration Set with 100 mL Bag and Cassette, 56'', Volume 3 mL, 100 mL Bag with Fill Port, Cassette with AutoClamp Device, Slide Clamp, Male Luer Lock Adapter b) Sabraset 56050 250 Administration Set with 2500 mL Bag and Cassette, 53'', Volume 3 mL, 250 mL Bag with Fill Port, Cassette with AutoClamp Device, Slide Clamp, Male Luer Lock Adapter
- Recall number
- Z-0386-05
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Sabraset Administration Sets with AutoClamp; for use with model 6060 multi-therapy pumps only; a prescription, sterile, non-pyrogenic fluid pathway; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., the following sets were made in Mexico: a) Sabraset 56050 100 Administration Set with 100 mL Bag and Cassette, 56'', Volume 3 mL, 100 mL Bag with Fill Port, Cassette with AutoClamp Device, Slide Clamp, Male Luer Lock Adapter b) Sabraset 56050 250 Administration Set with 2500 mL Bag and Cassette, 53'', Volume 3 mL, 250 mL Bag with Fill Port, Cassette with AutoClamp Device, Slide Clamp, Male Luer Lock Adapter
- Status
- Terminated
- Initiated
- 2004-11-19
- Posted
- 2005-01-08
- Terminated
- 2006-06-28
- Reason
- A missing platen assembly in the cassette of the solution sets results in the infusion pump being unable to occlude the tubing of the set. This can lead to free-flow of the solution and a possible overdose of medication to the patient.
- Root cause
- Other
- Action
- Urgent Product Recall letters dated 11/19/04 were sent to the direct accounts on the same date via UPS ground. The accounts were informed that free-flow situations were associated with the listed product codes and lot numbers due to a missing platen assembly in the cassette of the solution sets, which could result in an overdose of medication to the patient. The accounts were requested to stop use of the affected lot numbers and product codes and return the sets to Baxter c/o NNC Group, 2670 Executive Dr., Indianapolis, IN 46241, following the enclosed instructions. Any questions concerning the return process were directed to NCC at 1-866-300-5085. Any medical or technical questions were directed to Baxter at 1-800-422-9837.
- Quantity
- 1,355,820 sets
- Distribution
- Nationwide, including Puerto Rico, and internationally to Australia, Belgium, Canada, Hong Kong, Israel, Japan, Singapore, Sweden and Switzerland
- Lot, serial or model codes
- Product codes 560500-100 and 560500-250, all lot numbers other than lots produced within the range of 417519 to 499999, or those with an adhesive circular colored sticker (Avery dot) at the top of the individual package on the Tyvek side or on the shipper label. Any lot numbers from the following obsolete Sabraset product codes are also being recalled: 5600PM, 560100, 560101, 560110, 560111, 560112, 560112-GEL, 560112-L, 560113, 560115, 560116, 560122-L, 561101, 567100, 567100-L, 567112-L, 567122-L, 569100, 569111, 569122, 560200-100, 560522-100
- 510(k) numbers
- ["K002679"]
- PMA numbers
- not on file
- Product code
- FPA
- Firm FEI
- 1416980
- Firm city
- Round Lake
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28