Walter Lorenz Surgical, Inc.: Product is 35mm Flex Tube***NON-STERILE PRODUCT***CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN." W.LORENZ SURGICAL***WALTER LORENZ SURGICAL, INC. 1520 TRADEPORT DRIVE , REPLY TO POST OFFICE BOX 18009, JACKSONVILLE, FLORIDA 32229-8009 USA, 1-800-874-7711. Part # SP-1328-01 Lot 968720 Qty 34 10/6/2005. Sold for use with: Part # SP-1598-13 Lot # 041950 Qty 10. Part # SP-1597-13 Lot # 066720 Qty 9. Part # SP-1598-13 Lot 050140 Qty 15. Product Usage: Distraction devices are for use in bone stabilization and elongation (lengthening) when correction of oral, cranial, mandibular and maxiollofacial deficiencies or post traumatic defects require gradual bone distraction. Each implantable distraction device has a titanium alloy drive screw mechanism and LactoSorb resorbable correction plates, which are implanted by fixation with LactoSorb bone screws.

Recall number
Z-0385-2012
Recalling firm
Walter Lorenz Surgical, Inc.
Product
Product is 35mm Flex Tube***NON-STERILE PRODUCT***CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN." W.LORENZ SURGICAL***WALTER LORENZ SURGICAL, INC. 1520 TRADEPORT DRIVE , REPLY TO POST OFFICE BOX 18009, JACKSONVILLE, FLORIDA 32229-8009 USA, 1-800-874-7711. Part # SP-1328-01 Lot 968720 Qty 34 10/6/2005. Sold for use with: Part # SP-1598-13 Lot # 041950 Qty 10. Part # SP-1597-13 Lot # 066720 Qty 9. Part # SP-1598-13 Lot 050140 Qty 15. Product Usage: Distraction devices are for use in bone stabilization and elongation (lengthening) when correction of oral, cranial, mandibular and maxiollofacial deficiencies or post traumatic defects require gradual bone distraction. Each implantable distraction device has a titanium alloy drive screw mechanism and LactoSorb resorbable correction plates, which are implanted by fixation with LactoSorb bone screws.
Status
Terminated
Initiated
2006-02-17
Posted
2012-01-11
Terminated
2011-12-12
Reason
Lorenz Engineer discovered flex tube could break if bent too much.
Root cause
Device Design
Action
W. Lorenz Surgical notified consignees of recall by an Urgent Advisory Notice letter. Consignees were requested to acknowledge receipt of recall and provide product disposition. Initial letter was sent on March 2, 2006, March 15, 2006, follow-up letters sent on April 4, 2006 and May 15, 2006. For additional information call (904) 741-4400 or fax (904) 741-4500.
Quantity
34 pieces
Distribution
Worldwide Distribution - USA (nationwide) including the states of: GA and TX and the country of Japan.
Lot, serial or model codes
Lot # 968720
510(k) numbers
["K002083"]
PMA numbers
not on file
Product code
JEY
Firm FEI
1032347
Firm city
Jacksonville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register