FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter labeled 6060 Solution Sets with AutoClamp; For use with model 6060 multi-therapy pumps only; a prescription, sterile, non-pyrogenic fluid pathway; Manufactured by an affiliate of Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., the following sets were made in Costa Rica: a) Baxter Solution Set, 76'' (1.9 m), Volume 3.7 mL, Cassette with AutoClamp Device, Male Luer Lock Adapter, product code 2M9856 and 2M9856K; b) Baxter Solution Set, 109'' (2.8 m), Volume 4.8 mL, Cassette with AutoClamp Device, Male Luer Lock Adapter, product code 2M9857; c) Baxter Solution Set, 112'' (2.9 m), Volume 7.9 mL, Cassette with AutoClamp Device, 1.2 Micron Downstream Filter, Male Luer Lock Adapter, product code 2M9858 and 2M9858K; d) Baxter Solution Set, 112'' (2.9 m), Volume 7.9 mL, Cassette with AutoClamp Device, 0.22 Micron Downstream Filter, Male Luer Lock Adapter, product code 2M9859; e) Baxter Solution Set, 76'' (1.9 m), Volume 3.7 mL, Non-DEHP Tubing, Cassette with AutoClamp Device, Male Luer Lock Adapter, product code 2M9860; f) Baxter Solution Set, 112'' (7.9 m), Volume 3.7 mL, Non-DEHP Tubing, Cassette with AutoClamp Device, 1.2 Micron Downstream Filter, Male Luer Lock Adapter, product code 2M9861; g) Baxter Solution Set, 112'' (7.9 m), Volume 3.7 mL, Non-DEHP Tubing, Cassette with AutoClamp Device, 0.22 Micron Downstream Filter, Male Luer Lock Adapter, product code 2M9862; h) Baxter Solution Set, 111'' (2.9 m), Volume 7.4 mL, Cassette with AutoClamp Device, 1.2 Micron Downstream Filter, Male Luer Lock Adapters, product code 2M9874K; i) Baxter Solution Set, 75'' (1.9 m), Volume 3.1 mL, Cassette with AutoClamp Device, Male Luer Lock Adapters, product code 2M9875K; j) Baxter Epidural Set, 115'' (2.9 m), Volume 5.0 mL, Cassette with AutoClamp Device, Male Luer Lock Adapter, product code 2L9003; k) Baxter Epidural Set, 112'' (2.9 m), Volume 7.9 mL, Cassette with AutoClamp Device, 0.22 Micron Downstream Filter, Male Luer Lock Adapter, product code 2L9004; l) Baxter Epidural Set, 92'' (2.3 m), Volume 6.7 mL, Cassette with AutoClamp Device, 0.22 Micron Downstream Filter, Male Luer Lock Adapters, product code 2L9005; m) Baxter Solution Set with Spike, 88'' (2.2 m), Volume 4.1 mL, Non-DEHP Tubing, Cassette with AutoClamp Device, Male Luer Lock Adapter, product code 2L9006;

Recall number
Z-0385-05
Recalling firm
Baxter Healthcare Corp.
Product
Baxter labeled 6060 Solution Sets with AutoClamp; For use with model 6060 multi-therapy pumps only; a prescription, sterile, non-pyrogenic fluid pathway; Manufactured by an affiliate of Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., the following sets were made in Costa Rica: a) Baxter Solution Set, 76'' (1.9 m), Volume 3.7 mL, Cassette with AutoClamp Device, Male Luer Lock Adapter, product code 2M9856 and 2M9856K; b) Baxter Solution Set, 109'' (2.8 m), Volume 4.8 mL, Cassette with AutoClamp Device, Male Luer Lock Adapter, product code 2M9857; c) Baxter Solution Set, 112'' (2.9 m), Volume 7.9 mL, Cassette with AutoClamp Device, 1.2 Micron Downstream Filter, Male Luer Lock Adapter, product code 2M9858 and 2M9858K; d) Baxter Solution Set, 112'' (2.9 m), Volume 7.9 mL, Cassette with AutoClamp Device, 0.22 Micron Downstream Filter, Male Luer Lock Adapter, product code 2M9859; e) Baxter Solution Set, 76'' (1.9 m), Volume 3.7 mL, Non-DEHP Tubing, Cassette with AutoClamp Device, Male Luer Lock Adapter, product code 2M9860; f) Baxter Solution Set, 112'' (7.9 m), Volume 3.7 mL, Non-DEHP Tubing, Cassette with AutoClamp Device, 1.2 Micron Downstream Filter, Male Luer Lock Adapter, product code 2M9861; g) Baxter Solution Set, 112'' (7.9 m), Volume 3.7 mL, Non-DEHP Tubing, Cassette with AutoClamp Device, 0.22 Micron Downstream Filter, Male Luer Lock Adapter, product code 2M9862; h) Baxter Solution Set, 111'' (2.9 m), Volume 7.4 mL, Cassette with AutoClamp Device, 1.2 Micron Downstream Filter, Male Luer Lock Adapters, product code 2M9874K; i) Baxter Solution Set, 75'' (1.9 m), Volume 3.1 mL, Cassette with AutoClamp Device, Male Luer Lock Adapters, product code 2M9875K; j) Baxter Epidural Set, 115'' (2.9 m), Volume 5.0 mL, Cassette with AutoClamp Device, Male Luer Lock Adapter, product code 2L9003; k) Baxter Epidural Set, 112'' (2.9 m), Volume 7.9 mL, Cassette with AutoClamp Device, 0.22 Micron Downstream Filter, Male Luer Lock Adapter, product code 2L9004; l) Baxter Epidural Set, 92'' (2.3 m), Volume 6.7 mL, Cassette with AutoClamp Device, 0.22 Micron Downstream Filter, Male Luer Lock Adapters, product code 2L9005; m) Baxter Solution Set with Spike, 88'' (2.2 m), Volume 4.1 mL, Non-DEHP Tubing, Cassette with AutoClamp Device, Male Luer Lock Adapter, product code 2L9006;
Status
Terminated
Initiated
2004-11-19
Posted
2005-01-08
Terminated
2006-06-28
Reason
A missing platen assembly in the cassette of the solution sets results in the infusion pump being unable to occlude the tubing of the set. This can lead to free-flow of the solution and a possible overdose of medication to the patient.
Root cause
Other
Action
Urgent Product Recall letters dated 11/19/04 were sent to the direct accounts on the same date via UPS ground. The accounts were informed that free-flow situations were associated with the listed product codes and lot numbers due to a missing platen assembly in the cassette of the solution sets, which could result in an overdose of medication to the patient. The accounts were requested to stop use of the affected lot numbers and product codes and return the sets to Baxter c/o NNC Group, 2670 Executive Dr., Indianapolis, IN 46241, following the enclosed instructions. Any questions concerning the return process were directed to NCC at 1-866-300-5085. Any medical or technical questions were directed to Baxter at 1-800-422-9837.
Quantity
3,063,359 sets
Distribution
Nationwide, including Puerto Rico, and internationally to Australia, Belgium, Canada, Hong Kong, Israel, Japan, Singapore, Sweden and Switzerland
Lot, serial or model codes
Product codes 2M9859, 2M9856, 2M9856K, 2M9857, 2M9858, 2M9858K, 2M9860, 2M9861, 2M9862, 2M9874K, 2M9875K, 2L9003, 2L9004, 2L9005, 2L9006; all sets with lot numbers below R03K25200.  The lot number R03XXXXXX (format RYYMDDNNC) is broken down as follows: R = manufacturing plant location; YY = numeric year (01, 02, 03); M = alpha month, A = January through L = December; DD = numeric day of month; NN = numeric batch number for the day, 01 - 99; C = numeric check digit generated by computer.
510(k) numbers
["K002679"]
PMA numbers
not on file
Product code
FPA
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register