Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes USA HQ, Inc.: Synthes Matrix Threaded Persuader Deformity is a comprehensive thoracolumbar pedical screw system designed to provide flexibility biomechanical performance and a solution to complex posterior pathological challenges.

Recall number
Z-0384-2014
Recalling firm
Synthes USA HQ, Inc.
Product
Synthes Matrix Threaded Persuader Deformity is a comprehensive thoracolumbar pedical screw system designed to provide flexibility biomechanical performance and a solution to complex posterior pathological challenges.
Status
Terminated
Initiated
2013-07-31
Posted
2013-11-22
Terminated
2015-08-31
Reason
Synthes became aware that during a procedure using the Matrix Threaded Persuader, the Reduction Insert malfunctioned causing pieces to break off. Once there is a breakage to the Reduction Insert it will not allow the instrument to function as required.
Root cause
Labeling design
Action
The firm, Synthes, sent an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter dated July 31, 2013 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to review your inventory and immediately remove the affected lots from stock; If you do have the affected devices, contact Synthes at 1-800-479-6329 to obtain the RA# and return the product to with completed form to Credit/Returns, Synthes, 1101 Synthes Avenue, Monument, CO 80132; complete and sign the Return Receipt Requested form via fax at 1-888-731-7955 or email to: FieldAction@synthes.com, even if you do not have the affected product. If you have questions, please call 610-719-5450.
Quantity
634
Distribution
Worldwide distribution: US (nationwide) and countries of Canada and Switzerland.
Lot, serial or model codes
Item Lot Number Number 03.632.008 4833901 03.632.008 6284014 03.632.008 6284015 03.632.008 6377412 03.632.008 6390270 03.632.008 6390308 03.632.008 6390309 03.632.008 6445711 03.632.008 6455761 03.632.008 6455786 03.632.008 6530998 03.632.008 6548747 03.632.008 6558995 03.632.008 6559002 03.632.008 6583188 03.632.008 6583189 03.632.008 6603334 03.632.008 6603335 03.632.008 6624550 03.632.008 6676663 03.632.008 6703414 03.632.008 6706733 03.632.008 6718634 03.632.008 6718635 03.632.008 4833901 03.632.008 6445711 03.632.008 6455761 03.632.008 6455786 03.632.008 6530998 03.632.008 6548747 03.632.008 6603334 03.632.008 6603335 03.632.008 6676663 03.632.008 6703414 03.632.008 6706733 03.632.008 6718634 03.632.008 6718635
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3005180112
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register