Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens, Mevatron M2 Primus, Part No.: 19 40 035, Mid-Energy Medical Charge particle radiation therapy system, with pre-R2.0 Optivue Flat Panel Positioners, Siemens Medical, Concord, CA 94520

Recall number
Z-0384-2008
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens, Mevatron M2 Primus, Part No.: 19 40 035, Mid-Energy Medical Charge particle radiation therapy system, with pre-R2.0 Optivue Flat Panel Positioners, Siemens Medical, Concord, CA 94520
Status
Terminated
Initiated
2007-03-23
Posted
2007-12-01
Terminated
2011-02-11
Reason
Product's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force.
Root cause
Device Design
Action
The firm has issued a Customer Safety Advisory Notice dated 12/6/2006, to all sites that have the pre-R2.0 OPTIVUE Flat Panel Positioners installed. Users will have a new microswitch-based collision detection system, installed free of charge.
Quantity
9 units.
Distribution
Worldwide Distribution- USA including MA, ME, NJ, NY, PA, MD, WV, NC, SC, GA, FL, AL, TN, KY, OH, IN, MI, IA, WI, SD, ND, MT, IL, NE, LA, OK, TX, WY, NV, CA, AK AND AR, and countries of Serbia, Australia, Austria, Belgium, Canada, Denmark, El Salvador, France, Germany, Hungary, India, Ireland, Italy, Japan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Norway, China, Philippines, Poland, Portugal, Qatar, Republic of Korea, Russia, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Turkey, Ukraine, UK.
Lot, serial or model codes
Model number 1940035, Serial Numbers 3736 3756 3491 3849 3885 4087 70-4249 70-4286 70-4323
510(k) numbers
not on file
PMA numbers
not on file
Product code
IYE
Firm FEI
2937457
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register