Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Navigation, Inc.: Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, Model: 9733763, and StealthStation Cranial, Model: 9735585

Recall number
Z-0383-2022
Recalling firm
Medtronic Navigation, Inc.
Product
Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, Model: 9733763, and StealthStation Cranial, Model: 9735585
Status
Open, Classified
Initiated
2021-11-11
Posted
not on file
Terminated
not on file
Reason
Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, tissue injury.
Root cause
Software design
Action
On 11/11/21, correction notices were distributed to healthcare professionals. The recalling firm will provide a warning and instructional placard to be applied to impacted systems. Customers with questions are encouraged to call Technical Services for help at: 1-888-826-5603. Customers are asked to complete and return the confirmation form and maintain a copy of this correction notice. Until a software update is available, ensure you are following the instructions below to prevent the issue from occurring: Firm recommends NOT turning off Guidance View after locking the biopsy trajectory in the Cranial software. After locking trajectory for biopsy needle navigation, always ensure Guidance View remains an active view in at least one screen layout quadrant. If a discrepancy is detected between the biopsy depth gauge graphic and the other 2D information: 1) Obstruct the camera field of view of the biopsy needle or reference frame to cause red status. 2) Return to green status by no longer obstructing camera field of view of the biopsy needle and reference frame. 3) Use the Cycle Views icon to refresh guidance view and confirm that the biopsy depth gauge graphic matches the distance to target values and the position to target information provided by the 2D crosshairs. 4) Visually confirm accuracy before proceeding with navigation. Always Visually confirm navigational accuracy and confirm that the biopsy depth gauge graphic matches the distance to target values and matches the position to target information provided by the 2D crosshairs, which represent the tip position of the navigated instrument. Use the Biopsy Needle Mechanical Depth Stop. If system navigation seems inaccurate and steps to restore accuracy are unsuccessful, abort use of the system.
Quantity
2587
Distribution
US: MN, FL, MD, IL, AK, NE, NH, PA, ND, AR, CA, WI, NJ, GA, SD, AZ, KY, AL, MS, MO, TX, MI, MT, OH, NY, VA, MA, CT, TN, WV, WA, NC, CO, NV, ME, LA, DE, NM, IN, UT, DC, OR, SC, OK, IA, HI, ID, KS, WY, MN, PR OUS: AU, BN, GU, JP, KR, MY, NZ, SG, TH, VN, MM, CA, CZ, RU, AL, CY, DE, ES, FR, GB, HU, IL, IT, NL, PL, RO, CN, HK, TW, IN, LK, NP, AR, BB, BR, CL, CO, CR, DO, EC, MX, PE, PR, VE, NI, UY, AE, BH, DZ, EG, IQ, JO, KZ, KE, SA, LB, LY, MA, NA, PK, QA, ZA, TR, AT, BE, CH, DK, FI, GF, GR, IE, LV, LT, LU, PT, SE, SK, UK, BA, BG, CI, HR, RS, SI
Lot, serial or model codes
9733763, Version: 2.2.8; 9735585, Versions: 3.1.1, 3.1.2, 3.1.3
510(k) numbers
["K150216","K190672"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000517638
Firm city
Louisville
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register