Skip to the content

FDA Medical Device Recalls (openFDA)

TEI Biosciences: PriMatrix is an acellular dermal tissue matrix derived from fetal bovine dermis. The device is supplied sterile in a variety of sizes to be trimmed by the surgeon to meet the individual patient's needs.

Recall number
Z-0383-2018
Recalling firm
TEI Biosciences
Product
PriMatrix is an acellular dermal tissue matrix derived from fetal bovine dermis. The device is supplied sterile in a variety of sizes to be trimmed by the surgeon to meet the individual patient's needs.
Status
Terminated
Initiated
2017-06-15
Posted
not on file
Terminated
2018-03-15
Reason
During Final Assembly Operation, it was noticed that some of the PriMatrix Instructions for Use (IFUs) from Lot RM1608011 Part 520-002v10 contained extra pages and/or missing pages: " 22 IFUs contained duplicate pages 15 - 26 & 51 - 62 and were also missing pages 27 - 50 " 3 IFUs were missing pages 27 - 50 " 1 IFU was missing pages 1 - 2 & 75 - 76 " 1 IFU was missing pages 15 - 62 While pages 27 50, 75 76, and 15 62 contain instructions written in a language other than English, pages 1 2 contain information regarding the indications for use, contraindications, device description, warning and precautions, and potential complications of the device.
Root cause
Error in labeling
Action
Integra LifeSciences Corporation notified (all U.S.) customers via Letter/Medical Device Correction Field Safety Notice dated June 15, 2017. Letter delivered by courier service, facsimile or email July 15, 2017.
Quantity
4081 units
Distribution
U.S.
Lot, serial or model codes
Product Model Number Product Lot Number Product Description Product Model Number Product Lot Number Product Description 607-004-880 1508019 8 x 8 cm 607-005-014 1702023 14mm Disc 607-005-880 1509051 8 x 8 cm 607-005-018 1702024 18mm Disc 607-005-440 1512021 4 x 4 cm 607-005-220 1702025 2 x 2 cm 607-005-812 1512022 8 x 12 cm 607-005-014 1702026 14mm Disc 607-005-440 1603044 4 x 4 cm 607-005-018 1702027 18mm Disc 607-005-660 1606038 6 x 6 cm 607-005-220 1702028 2 x 2 cm 607-001-112 1611044 10 x 12 cm 607-005-014 1702029 14mm Disc 607-001-812 1611045 8 x12 cm 607-005-018 1702030 18mm Disc 607-001-225 1612013 20 x 25 cm 607-005-220 1702031 2 x 2 cm 607-001-009 1612018 0.2 x 26.5 cm 607-005-014 1702032 14mm Disc 607-004-440 1701034 4 x 4 cm 607-005-014 1702033 14mm Disc 607-001-440 1701036 4 x 4 cm 607-005-014 1702034 14mm Disc 607-004-660 1701037 6 x 6 cm 607-005-014 1703007 14mm Disc 607-001-660 1701044 6 x 6 cm 607-005-018 1703008 18mm Disc   The affected lots are manufactured from February 2017 to May 2017.   The affected lots will expire in 2020 to 2021.
510(k) numbers
["K153690"]
PMA numbers
not on file
Product code
KGN
Firm FEI
3004170064
Firm city
Boston
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register