Skip to the content

FDA Medical Device Recalls (openFDA)

Accuray Incorporated: CyberKnife Robotic Radiosurgery System: Lung Optimized Treatment option. The CyberKnife treatment is indicated for the treatment planning and image guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Recall number
Z-0382-2012
Recalling firm
Accuray Incorporated
Product
CyberKnife Robotic Radiosurgery System: Lung Optimized Treatment option. The CyberKnife treatment is indicated for the treatment planning and image guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Status
Terminated
Initiated
2011-11-07
Posted
2012-01-11
Terminated
2012-02-27
Reason
User Facility reported an anomaly where inhale and exhale CT pairs used for treatment planning did not represent the same magnitude of respiration that was displayed during treatment delivery.
Root cause
Labeling design
Action
Accuray sent a Urgent Medical Device Correction letter dated November 7, 2011, via Federal Express to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The information provided in this letter will be incorporated in user manuals shipped with the CyberKnife system. If you have any questions or concerns regarding this issue, please do not hesitate to contact Accuray Customer Support at 1. (877) 668-8667 ( USA) 1. (408) 716-4700 ( non USA )
Quantity
30
Distribution
Worldwide Distribution -- USA ( nationwide ) including the states of FL, TN, IL, NJ, CO, PA, MD, CA, MS, MN, SC, WA, OK, GA, AZ, and TX., and the countries of France and United Kingdom.
Lot, serial or model codes
Model number: PN 032384
510(k) numbers
["K091999"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1000123585
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register